MINI6600
TM
Mobile Digital C-Arm Service Manual - Periodic Maintenance
4. Measure the line voltage at the wall receptacle while making a fluoroscopic exposure at 98
µ
A @ 75kV
(Max load voltage).
V1=_____
5. Use the following formula to calculate the percentage of line voltage regulation and record the results
below.
Line Regulation = 100 X (Vn - V1)
V1
where:
Vn = No load voltage
V1 = Max load voltage
Regulation = _________%
6. If the results above do not fall within the range of 0 to 15 percent, the hospital should be informed that
an in-house wiring problem exists that should be corrected as soon as possible. Note this on the FSR.
(This problem results in a violation of a provision of the Federal Performance Standards 21 CFR
1020.30)