SW22
5
User
manual
P1630830G
2 – GENERAL FUNCIONS
2.1 – Tare Function
The Tare function allows to clear the weight of a container on the plate of the scale.
Pressing key
the display will be zeroed and the Net Led will turn on.
The Tare operation can be carried out several times ; the maximum tare value is equal to the
scale capacity.
If the Tare is locked, unloading the scale, the display will show a negative value while if the
Tare is unlocked, the tare value will be cleared.
2.2 – Numeric Tare
The numeric Tare function allows to set a Tare value; it’s possible, starting from a full
container, to spin-off the Tare value and display the Net weight value
Input the numeric value of the tare ( including dot with key and confirm with
.
The entered value is automatically rounded to the division currently active.
If you enter a value greater than full scale, the value is not accepted.
2.3 – Lock Tare
It is possible to lock the Tare value stored so that it will be stored also unloading the scale.
Press the keys
and
.
The display will show “T BL” ; starting from now, the tare value is locked.
To unlock the Tare value , Press again the keys
and
; the display will show “T SBL”
starting from now, the tare value is unlocked.
2.4 – Archive Tare creation
The operator has the ability to create an archive of 9 tare values that can be called directly by
entering their identification code.
To access the tares, press keys in sequence
and
.
The display is ready for entering the identification code of the tare you want to enter or change (1
to 9) displaying "T - 1"; the selection is made with the
▲
and
▼
and confirm with
.
The display will show the tare value actually in memory at that location or 00000.0; enter the tare
weight and confirm with
; The display will now show "MPL 01" corresponding to the
multiplication factor to be applied to just set tare; entering a value other than 1, each time you
select this tare automatically be multiplied by the factor set.
To exit press the tare database management press key
.
2.5 – Tare recall from archive
To recall a previously stored in the archive of Tare press the number key from 1 to 9 corresponding
to it followed by
; the display will show the value corresponding tare. Then confirm with
If the memory location contains a tare value equal to 0, pressing the button will have no effect.
2.6 – Clear a Tare value
To clear a stored tare value, Press
for 3 sec. .
2.7 – Multiplicative Tare factor
The operator has the possibility of introducing a multiplicative factor that will be applied to the
value of the tare memory.
To access this feature, press the button
for 3 sec.
The display will show "No xx" where xx is the numerical value of the multiplicative factor active
at the moment; to change this value, enter the desired number from the keypad and confirm
with
.
The values are between 1 and 99.
From this moment the tare value assets, will be multiplied by this factor.
To cancel the effect of multiplication, proceed as described above, and set a value of 1.