9 Preparing for Capturing Images
Instruction Manual ImageCam® 3 (G/69200/XXXX/EN 0120 Rev01)
19 / 68
9.3.4
Editing Pre-Set Exposure Programmes, If Necessary
To achieve the best possible image quality, OCULUS has pre-set the
camera parameters for common image capture situations and has saved
them as exposure programmes. You can edit these settings,
.
You can then save the edited settings as your own exposure programme,
.
The following exposure programmes have been pre-set:
☞
If you are not satisfied with the image quality, you can optimise the
images using the pre-settings; also see
.
Diffuse
illumination
Optical
section
Fluo
images
Fundus
images
Slit lamp
Medium bright
Very bright
Very bright
Minimally bright
Heat protection
filter
a
Swung in
Swung in
Swung in
Swung in
Diffuser
Swung forward
Blue filter
In the illumination
beam path
Yellow filter
In the observation
beam path
Slit width
As widely as
possible
As small as
possible
As widely as
possible
To match
pupil diameter
Magnification
10 - 12.5 x
10 - 12.5 x
10 - 12.5 x
20 x
Slit length
Maximum
Maximum
Maximum
To match
pupil diameter
Beam splitter
Aperture
medium to open
Aperture
open
Aperture
open
Aperture
open
Exposure time
20 ms
30 ms
35 ms
38 ms
Gain
0%
35%
20%
62%
Gamma
1.30
1.0
1.2
1.3
a.If present
Summary of Contents for ImageCam 3
Page 1: ...OCULUS ImageCam 3 INSTRUCTION MANUAL Slit Lamp Documentation System...
Page 6: ...v v Instruction Manual ImageCam 3 G 69200 XXXX EN 0120 Rev01...
Page 73: ...19 Appendices Instruction Manual ImageCam 3 G 69200 XXXX EN 0120 Rev01 67 68...
Page 74: ...19 Appendices 68 68 Instruction Manual ImageCam 3 G 69200 XXXX EN 0120 Rev01...
Page 75: ......