52
05.04.2011 AD W9002 N
The machine calibration is fi nished.
051
6.2
Calibration of the gauge arms
Touching the button “GAUGE ARMS”
starts the calibration of the inner and
outer
gauge
arm.
053
Attach the inner gauge arm to the fl ange and cali-
bration weights as show on the screen and wait for an
acoustic signal.
054
Move the arm as shown on the screen to the front and
wait for an acoustic signal.
Only BM 45 Touch and BM 55 Touch
055
Attach the outer gauge arm as shown on the screen to
the tip of the calibration weight and wait for an acoustic
signal.
056
Bring both gauge arms to their home position and wait
for an acoustic signal.