Insert the spacing gage “0,6” lengthwise between the scale tape surface and the
scanning head�
• Push the scanning head against the spacing gage and tighten the two mounting screws
alternatingly and evenly (tightening torque = 1 Nm). The spacing gage should barely be
movable�
• Check the parallelism of the scanning head to the scale tape� Push the spacing gage ca�
5 mm to the left and right between the scanning head and scale tape, checking whether
the clearance is the same�
•
Remove the spacing gage cautiously. If it is difficult to remove the spacing gage, loosen
the scanning head and repeat the mounting procedure� Otherwise the surface of the
sensor module can be damaged!
9
!
• Position the encoder cable including the strain relief near the scanning head�
• Connect the encoder cable to the evaluation electronics and pay attention to the pin
assignment. The evaluation electronics must be switched off!
• After connecting the encoder cable switch the evaluation electronics on and test the
functionality�
• If there are errors in the function or during measurement, please read chapter 12
“Troubleshooting” first. If these information are not helpful for your problem, please
contact the technical support of NUMERIK JENA or an authorized representation in
your country�