Appendix A1
MC-II
Plus
WP-PM Flow Analyzer
May 2005
A1-1
Appendix A1 – Installation – Panel Mount MC-II Plus
General
The Panel Mount MC-II
Plus
Flow Analyzer is available in three configurations, ¼
DIN, Double Wide Single Readout and Double Wide Two Readout.
The ¼ DIN unit mounts in the DIN standard ¼ DIN square opening (3.62” X 3.62”).
Due to variations in mounting situations some mounting fasteners are to be supplied
by the installer. Studs (#6-32 X ¾”) and nuts (#6-32) with integral lockwasher are
supplied for mounting the instrument. Mounting dimensions are shown on page A1-
7.
The Double Wide Single Readout unit fits in an industry size rack and is especially
suited for replacement of Double Wide NuFlo Model LO-II Flow Totalizers as well as
new installations. Mounting dimensions are shown on page A1-8.
The Double Wide Two Readout unit fits in the same size rack as the Double Wide
Single Readout unit. This instrument provides two readouts in the same space as
one readout as supplied in the above unit. It is especially suited for replacing one
Double wide NuFlo Model LO-II Flow Totalizer with two Panel Mount MC-II
Plus
Flow Analyzers. It is also suited for replacing two standard width NuFlo Model LO-II
Flow Totalizers in the same space. Mounting dimensions are shown on page A1-9.
Input / Output Features
There are five input/output features available for the Panel Mount MC-II
Plus
Flow
Analyzer. Each feature is covered individually in the following sections with wiring
diagrams.
Pulse Output
The Pulse Output is provided in the form of an optically isolated open collector
transistor circuit. It can be used in conjunction with any other feature on the Panel
Mount MC-II
Plus
Flow Analyzer. A two-conductor cable from the Panel Mount MC-
II
Plus
Flow Analyzer to the remote location is required with a 5 to 30 VDC power
supply and suitable device for reading the open collector pulse output of the Panel
Mount MC-II
Plus
Flow Analyzer. The maximum current rating of the pulse output
circuit is 40 mA @ 30 VDC. The Pulse Output Wiring Diagram is located on page
A1-11.
The section in the manual,
Configuring the Pulse Output
, provides information
regarding the setup of the pulse output feature.
Summary of Contents for MC-II Plus Panel Mount
Page 44: ...MC II Plus WP PM Flow Analyzer Appendix A1 A1 10 May 2005 Flow Meter Input Wiring...
Page 46: ...MC II Plus WP PM Flow Analyzer Appendix A1 A1 12 May 2005 External Power Wiring...
Page 47: ...Appendix A1 MC II Plus WP PM Flow Analyzer May 2005 A1 13 4 4 to 20mA Rate Output Wiring...
Page 52: ...MC II Plus WP PM Flow Analyzer Appendix A1 A1 18 May 2005...
Page 62: ...MC II Plus WP PM Flow Analyzer Appendix A2 A2 10 May 2005...
Page 72: ...MC II Plus WP PM Flow Analyzer Appendix C C 8 May 2005 NOTES...
Page 78: ...MC II Plus WP PM Flow Analyzer Appendix D D 6 May 2005 NOTES...
Page 92: ......