MC-II
Plus
WP-PM Flow Analyzer
May 2005
25
and 100. The 0.001 and 0.01 factors may not be used since the pulse output would
be set to increment faster than the volume display.
The valid selections for the Pulse Output Scale Factor are:
Volume Decimal Point
Valid Selections for Scale Factor
0.001
0.001
, 0.01, 0.1, 1, 10, 100
0.01
0.01
, 0.1, 1, 10, 100
0.1
0.1
, 1, 10, 100
1
1
, 10, 100
The Pulse Output Scale Factor will automatically change if the MC II
Plus
is returned
to the Run Mode without checking or changing the Pulse Output Scale Factor after
the volume decimal point is changed and the previous Pulse Output Scale Factor is
at a faster update rate than the volume. The scale factor will change to a value
equal to the volume decimal point setting.
For example, assume the volume decimal point is set to 0.00 and the Pulse Output
Scale Factor is set to 0.01. The Calibrate Mode is entered, the volume decimal point
is changed to 0.0, and the unit is returned to the Run Mode without checking or
changing the Pulse Output Scale Factor. The Pulse Output Scale Factor will
automatically change to 0.1. The value that the Pulse Output Scale Factor will
change to for the volume decimal point positions is shown in
bold
in the above table.
The Pulse Length Duration (Pulse Width)
– This factor determines the length of
each output pulse in milliseconds (ms). There are six user-selectable pulse lengths:
65 ms, 130 ms, 195 ms, 260 ms, 520 ms, and 1040 ms.
The steps required to configure the pulse output are:
1. Enter the Calibrate Mode by pressing the
ACCESS
key.
2. Accept the current settings by pressing
ENTER
until the
PULS.oUt
prompt
appears.
3. Press
INCR
to toggle the bottom line to
on
in order to enable the feature. If the
output is not used, press
INCR
until
oFF
is displayed on the bottom line. Press
ENTER
.
4. At the
PULS.div
prompt, press
INCR
to change the Pulse Output Scale Factor.
Press
ENTER
when the desired setting is displayed.
5. At the
PULS.Lng
prompt, press
INCR
to change the Pulse Length Duration.
Press
ENTER
when the desired setting is displayed.
6. At this point, the pulse output has been configured. Pressing
ACCESS
exits
the Calibrate Mode without making any further changes, or the remaining steps
may be implemented.
Summary of Contents for MC-II Plus Panel Mount
Page 44: ...MC II Plus WP PM Flow Analyzer Appendix A1 A1 10 May 2005 Flow Meter Input Wiring...
Page 46: ...MC II Plus WP PM Flow Analyzer Appendix A1 A1 12 May 2005 External Power Wiring...
Page 47: ...Appendix A1 MC II Plus WP PM Flow Analyzer May 2005 A1 13 4 4 to 20mA Rate Output Wiring...
Page 52: ...MC II Plus WP PM Flow Analyzer Appendix A1 A1 18 May 2005...
Page 62: ...MC II Plus WP PM Flow Analyzer Appendix A2 A2 10 May 2005...
Page 72: ...MC II Plus WP PM Flow Analyzer Appendix C C 8 May 2005 NOTES...
Page 78: ...MC II Plus WP PM Flow Analyzer Appendix D D 6 May 2005 NOTES...
Page 92: ......