
MC-II
Plus
WP-PM Flow Analyzer
May 2005
5
When the
Error
message is displayed, press any of the four keys on the keypad and
the first error will be shown on the upper line of the display. Press
INCR
or
STEP
to
check for additional errors. Each time
INCR
or
STEP
is pressed, the next error will
be displayed. After the last error is displayed, pressing
INCR
or
STEP
again will
display the first error again and the process will be repeated. It is recommended that
the
INCR
or
STEP
keys be pressed repeatedly to display all
Error
messages. The
ENTER
key may be pressed to return to the Run Mode or the
ACCESS
key may be
pressed to enter the Calibrate Mode to correct the errors as shown below in this
section.
If more than one
Error
condition exists, and one of these errors is corrected, the
next
Error
will be displayed. If all
Error
conditions cease to exist, the MC-II
Plus
Flow Analyzer will automatically return to the Run Mode.
There are four
Error
messages in the current version of firmware:
rAtE, PULS.oUt
,
4-20.oUt/Err Hi
and
4-20.oUt/Err Lo
.
The
rAtE
error message indicates a rate overflow. This means that the rate is in
units too large to be displayed on the LCD. Normally this error may be corrected by
entering the Calibrate Mode and changing the rate units of measure. These
changes are covered in the
Calibration
section of the manual.
The
PULS.oUt
error message indicates the pulses are accumulating faster than the
unit can output them. When pulses in excess of 255 have accumulated internally,
the MC-II
Plus
Flow Analyzer displays the
PULS.oUt
error message. The MC-II
Plus
Flow Analyzer can internally accumulate 65535 pulses, so the
Error
message
occurs long before pulses are lost. The MC-II
Plus
Flow Analyzer will continue to
output the accumulated pulses until the accumulated number drops to zero, even if
the flow rate stops. This ensures that no pulses will be lost unless the accumulated
value exceeds 65535. Normally this error may be corrected by entering the
Calibrate Mode and selecting a larger pulse output scale factor and/or a shorter
pulse width duration. These changes are covered in the
Configuring Pulse Output
section of the manual.
The
4-20.oUt/Err Hi
error message indicates the flow rate has exceeded the full-
scale calibrated flow rate setting to the point that the current output has exceeded 22
mA. The
Error
message appears on the lower line of the display only during the
time that the condition exists. When any key is pressed during the error condition,
the
4-20.oUt
message appears on the upper line of the display and the
Err Hi
message appears on the lower line of the display. This error may be caused by
excessive flow rate or the full scale flow rate calibration point being set too low for
normal operating conditions. Changing the full-scale flow rate calibration point is
covered in
Configuring the 4-20 mA Rate Output
section of the manual.
The
4-20.oUt/Err Lo
error message indicates the flow rate is below the calibrated
low flow rate setting. The output current will be 3.9 mA. The
Error
message
Summary of Contents for MC-II Plus Panel Mount
Page 44: ...MC II Plus WP PM Flow Analyzer Appendix A1 A1 10 May 2005 Flow Meter Input Wiring...
Page 46: ...MC II Plus WP PM Flow Analyzer Appendix A1 A1 12 May 2005 External Power Wiring...
Page 47: ...Appendix A1 MC II Plus WP PM Flow Analyzer May 2005 A1 13 4 4 to 20mA Rate Output Wiring...
Page 52: ...MC II Plus WP PM Flow Analyzer Appendix A1 A1 18 May 2005...
Page 62: ...MC II Plus WP PM Flow Analyzer Appendix A2 A2 10 May 2005...
Page 72: ...MC II Plus WP PM Flow Analyzer Appendix C C 8 May 2005 NOTES...
Page 78: ...MC II Plus WP PM Flow Analyzer Appendix D D 6 May 2005 NOTES...
Page 92: ......