
MC-II
Plus
WP-PM Flow Analyzer
18
May
2005
Press
STEP
to proceed to the next digit (thousands position). Press
INCR
until 6
is displayed.
Press
STEP
to proceed to the next digit (ten thousands position). Press
INCR
until 8 is displayed.
Press
STEP
to proceed to the next digit (one hundred thousands position).
Press
INCR
until 7 is displayed.
Press
ENTER
to confirm the entry of the divisor.
7. The upper line of the display will show
rAtE d.P
prompting for entry of the rate
multiplier decimal point. The selections for the rate multiplier decimal point are
0.0, 0.00, 0.000, 0.0000 and 0.00000 (factory default is 0.0). Press
INCR
until
0.00000 is displayed. Press
ENTER
.
8. The upper line of the display will show
rAtE.FAC
prompting for entry of the rate
multiplier (factory default is 1.00000). Enter the rate multiplier (.10986) in the
same manner as the divisor was entered in Step 6 and press
ENTER
.
9. The upper line of the display will show the prompt
rAtE.dLY
. The lower line of
the display will show the flow rate filter value in terms of the number of samples
required to reach 90% of the final value. The factory default setting is “nonE.”
The available settings are nonE, 5, 10, and 20. Press
INCR
until 10 is displayed.
Then, press
ENTER
. (See the section
Setting Flow Rate Filter
to change the flow
rate filter.)
10. The upper line of the display will show
inP.SEnS
prompting for entry of the input
sensitivity in millivolts. Since the input sensitivity (factory default of 20 mV), pulse
output (factory default is off), 4-20 mA output (factory default is off) and security
code (factory default is off) are to be left at factory default settings, press
ACCESS
to return to Run Mode. (Keep in mind that the unit for this example
was at factory default. Bypassing these steps in this case is acceptable. If the
MC-II
Plus
Flow Analyzer had been previously calibrated and the settings of
these functions unknown, they must be stepped through to ensure that they are
set as desired.)
Gas Measurement Using a Calculated Divisor and Rate Multiplier
Calculating the divisor and rate multiplier for gases is necessary when registering in
units other than MCF and MCF/D. The
USEr
volume and rate functions of the MC-II
Plus
Flow Analyzer are used in this case. Each MC-II
Plus
Flow Analyzer is shipped
with a label set containing commonly used rate and volume units of measure labels.
The appropriate label can be placed on the front panel on the right side of the LCD
viewing window. The part number for the label set is listed in Appendix B in the
Summary of Contents for MC-II Plus Panel Mount
Page 44: ...MC II Plus WP PM Flow Analyzer Appendix A1 A1 10 May 2005 Flow Meter Input Wiring...
Page 46: ...MC II Plus WP PM Flow Analyzer Appendix A1 A1 12 May 2005 External Power Wiring...
Page 47: ...Appendix A1 MC II Plus WP PM Flow Analyzer May 2005 A1 13 4 4 to 20mA Rate Output Wiring...
Page 52: ...MC II Plus WP PM Flow Analyzer Appendix A1 A1 18 May 2005...
Page 62: ...MC II Plus WP PM Flow Analyzer Appendix A2 A2 10 May 2005...
Page 72: ...MC II Plus WP PM Flow Analyzer Appendix C C 8 May 2005 NOTES...
Page 78: ...MC II Plus WP PM Flow Analyzer Appendix D D 6 May 2005 NOTES...
Page 92: ......