![NSK Ti-Max NLX nano Operation Manual Download Page 8](http://html1.mh-extra.com/html/nsk/ti-max-nlx-nano/ti-max-nlx-nano_operation-manual_1714619008.webp)
English
7
NOTICE
The displayed air pressure is only a guideline. Measure the correct air pressure with the air meter.
NOTICE
· Default value is set to "3.0".
· When the air pressure is not between "2.0" - "4.0", it is displayed "Ec", and will not be memorized.
NOTICE
· Default value is set to " 3.5".
· Adjustable between ".6" - "3.6".
9) Control the Foot Pedal of the delivery system, and maintain the air pressure between "2.0" - "4.0". Press M2 key until
it beep (more than 3 seconds) to memorize the upper limit of the air pressure.
0) Press both M2 key and Speed Adjustment Key( ) at the same time until it beep (more than 3 seconds), once again.
When it alarmed, this setting has completed. By setting Indicator by pressing Rotation Speed Key to " 40", and
now it is ready to use at the maximum speed, 40,000 min
-
(rpm).
6-2 Light Brightness Adjusting Function
(
Unit of Display = V
)
Brightness of the Fiber Optic can be adjusted from this function.
) Press M2 Key and Speed Adjustment Key( ) at the same time until it beep (more than 3 seconds).
2) With alarm sound, Indicator will change to " C". Pressing the Gear Ratio Select key can change the Indicator and Gear
Ratio Lamp (C C2 C3 C….) Set indicator in C3 (Light Brightness adjustment mode).Press FWD/REV Rotation
Key to proceed.
3) When Indicator changed to " 3.5"(Default value), press Speed Adjustment Key( ) and adjust the brightness.
4) Press M2 key until it beep (more than 3 seconds) to set the brightness.
5) Press M2 key and Speed Adjustment Key( ) at the same time for more than 3 seconds, once again. When it alarmed,
this setting has completed.
6-3 Program Function
It is possible to memorize settings (Speed, Gear Ratio, Forward/Reverse Direction)
After the setting, you can use your settings by pressing M, M2 Key.
) Set the values for Speed, Gear Ratio, Forward/Reverse Direction.
2) Press M Key or M2 Key until it beep (more than 3 seconds). When it alarmed, this setting has completed.
6) After setting lower limit, press Gear Ratio Select Key. Indicator will change to "C2 " (Upper air pressure limit
setting mode).
7) By pressing the FWD/REV Rotaion Key, Indicator shows " 0.0" and "Ec" mutually.
8) By stepping the Foot Pedal of the delivery system, Indicator shows the current air pressure.