4
CAUTION
Ř
Hold the spray can upright.
Ř
Be sure to hold the motor firmly to prevent slipping out of hand by the spray pressure when
lubricating.
6-2 Cleaning
Manual Cleaning (External Surface)
1) Remove dirt and debris on the motor surface using a brush under running water (at 38°C or less). Do
not use a wire brush.
2) Wipe off the moisture on the surface using absorbent cloth.
Automatic Cleaning (External and Internal Surfaces) (Only for FX205m M4/FX205m B2)
Use a thermo-disinfector in accordance with ISO 15883 (EN ISO 15883).
Refer to the thermo-disinfector manual.
6-3 Disinfection
Manual Disinfection (External Surface)
Wipe the surface with cloth soaked in disinfectant solution such as alcohol.
Automatic Disinfection (External and Internal Surfaces) (Only for FX205m M4/FX205m B2)
Use a thermo-disinfector in accordance with ISO 15883 (EN ISO 15883).
Refer to the thermo-disinfector manual.
6-4 Lubrication
NSK PANA SPRAY Plus / PANA SPRAY
Apply NSK PANA SPRAY Plus / PANA SPRAY every time after each use and/or before autoclaving.
1) Remove the motor from the hose.
2) Mount the arrow-head spray nozzle tip into the spray can port. (Fig. 7)
CAUTION
Ř
After washing with thermo-disinfector and prior to lubrication, dry the product until all internal
moisture is thoroughly removed. Thermo-disinfector moisture remaining inside the product could
reduce the effect of lubrication and could cause corrosion inside of the product.
NOTICE
Ř
NSK recommends the use of "Spray Mist Absorber" (Y900084) to prevent oil mist expelling out of
the motor.
3) Insert the Spray Nozzle into the Drive Air Port of the motor. Hold the motor and spray for approximately
2-3 seconds. Apply lubricant until it expels from the exhaust air part for at least 2 seconds. (Fig. 8)
Lubrication Oil
1) Remove the motor from the hose.
2) Apply 1-2 drops of lubrication oil into the Drive Air Port shown in the Fig. 8.
3) Connect the motor to the hose. Operate the motor for about 10 seconds.
6-5 Sterilization
Sterilize the product by autoclave sterilization. After each patient sterilize as noted below.
1) Insert into an autoclave pouch. Seal the pouch.
2) Autoclavable under the conditions below.
Autoclave for more than 20min. at 121°C, or 15min. at 132°C, or 3min. at 134°C.
3) The motor should remain in the autoclave pouch until required for use.