NOVUS AUTOMATION
17/28
* Check the
Table 10
for more information about the device configuration registers.
Table 08 -
NP785 Model: Configuration Registers Table
The registers 103 to 106, 123 to 126, 157 and 158 must be used by the user to enter the values of their respective parameters. If they are within
limits, the device will automatically pass these values to registers 108 to 111, 132 to 135, 159 and 160, which show the values considered during
the operation. In case of extrapolation of limits, this condition will be signaled in register 343 (HR_DIAGNOSE03).
For 32-bit data, the two registers that compose them must be read and/or written in order for the values to be updated.
133
HR_A1SPHI_H
-290,08 mpsi
-8,029 inH2O
-203,94 mmH2O
-2000,0 Pa
290,08 mpsi
8,029 inH2O
203,94 mmH2O
2000,0 Pa
mpsi:
2 decimal places
inH2O:
3 decimal
places
mmH2O:
2 decimal
places
Pa:
1 decimal place
134
HR_A1SPLO
Alarm Setpoint Low
RO
int_32
0
-20 mbar
-290,08 mpsi
-8,029 inH2O
-203,94 mmH2O
-2000,0 Pa
20 mbar
290,08 mpsi
8,029 inH2O
203,94 mmH2O
2000,0 Pa
mbar:
3 decimal places
mpsi:
2 decimal places
inH2O:
3 decimal
places
mmH2O:
2 decimal
places
Pa:
1 decimal place
135
HR_A1SPLO_H
137
HR_BAUD
Baud Rate*
RW
uint16
4
0
7
N.A.
138
HR_PRTY
Parity*
RW
uint16
0
0
2
N.A.
139
HR_ADDR
Slave address
RW
uint16
1
1
247
N.A.
146
HR_OFST_PRESS
Differential pressure
offset
RW
uint16
0
-4 mbar
-58,02 mpsi
-1,606 inH2O
-40,78 mmH2O
-400,0 Pa
4 mbar
58,02 mpsi
1,606 inH2O
40,78 mmH2O
400,0 Pa
mbar:
3 decimal places
mpsi:
2 decimal places
inH2O:
3 decimal
places
mmH2O:
2 decimal
places
Pa:
1 decimal place
152
HR_OUT1_FORCE_ENAB
Enables forcing output 1
RW
uint16
0
0
1
N.A.
153
HR_OUT1_FORCE_VAL
Forced value for output
1
RW
uint16
0
0,00 V
3,60 mA
10,00 V
21 mA
2
154
HR_A1_FORCE_ENAB
Enables forcing alarm 1
RW
uint16
0
0
1
N.A.
155
HR_A1_STATE
Changes forced status
on alarm
RW
uint16
0
0
1
N.A.
156
HR_FORCE_IN_PRESS
Enables differential
pressure forcing
RW
uint16
0
0
1
N.A.
157
HR_FORCE_PRESS_IN
Forced differential
pressure value input
RW
int_32
0
-20 mbar
-290,08 mpsi
-8,029 inH2O
-203,94 mmH2O
-2000,0 Pa
20 mbar
290,08 mpsi
8,029 inH2O
203,94 mmH2O
2000,0 Pa
mbar:
3 decimal places
mpsi:
2 decimal places
inH2O:
3 decimal
places
mmH2O:
2 decimal
places
Pa:
1 decimal place
158
HR_FORCE_PRESS_IN_H
159
HR_FORCE_PRESS
Forced differential
pressure valure
RO
int_32
0
-20 mbar
-290,08 mpsi
-8,029 inH2O
-203,94 mmH2O
-2000,0 Pa
20 mbar
290,08 mpsi
8,029 inH2O
203,94 mmH2O
2000,0 Pa
mbar:
3 decimal places
mpsi:
2 decimal places
inH2O:
3 decimal
places
mmH2O:
2 decimal
places
Pa:
1 decimal places
160
HR_FORCE_PRESS_H
161
HR_RESET_MIN_MAX
Reset all lows and
highs
RW
uint16
0
0
1
N.A.
166
HR_PRODUCT_TAG01
Equipment tag
RW
2 char
0
0x0000
0xFFFF
N.A.
167
HR_PRODUCT_TAG02
RW
2 char
0
0x0000
0xFFFF
N.A.
168
HR_PRODUCT_TAG03
RW
2 char
0
0x0000
0xFFFF
N.A.
169
HR_PRODUCT_TAG04
RW
2 char
0
0x0000
0xFFFF
N.A.
170
HR_PRODUCT_TAG05
RW
2 char
0
0x0000
0xFFFF
N.A.
171
HR_PRODUCT_TAG06
RW
2 char
0
0x0000
0xFFFF
N.A.
172
HR_PRODUCT_TAG07
RW
2 char
0
0x0000
0xFFFF
N.A.
173
HR_PRODUCT_TAG08
RW
2 char
0
0x0000
0xFFFF
N.A.
174
HR_PRODUCT_TAG09
RW
2 char
0
0x0000
0xFFFF
N.A.
175
HR_PRODUCT_TAG10
RW
2 char
0
0x0000
0xFFFF
N.A.
www.
GlobalTestSupply
.com
Find Quality Products Online at: