![North Atlantic 75DS2 Operation Manual Download Page 24](http://html.mh-extra.com/html/north-atlantic/75ds2/75ds2_operation-manual_1707616024.webp)
DLV Two/Four Channel (Module 5*)
75DS2 Operations Manual
North Atlantic Industries, Inc.
11/11/2011
Rev: 2011-11-11-1000
www.naii.com
Page 24 of 43
DLV BIT Test Enable
Set bit to enable associated Built-In Self Test D2 or D3.
The on-line (D2) Test -
Writing “1” to the D2 bit of the
DLV Test Enable Register
initiates status reporting of the
automatic background BIT testing that checks the output accuracy of each channel by comparing the measured
output position to the commanded position. The status bits will be set to indicate an accuracy (0.05º) problem and
the results can be read from DLV Status Registers within 2 seconds and if enabled, an interrupt will be generated
(See Interrupt Register). Writing a “0” deactivates the status reporting. The testing is totally transparent to the
user, requires no external programming, and has no effect on the standard operation of this card.
Note:
Outputs
must be ON and Excitation supplied for test to function. Card will write 55h (every 0.1 seconds) to the DLV Test
(D2) Verify Register when D2 is enabled. User can periodically clear to 00h and then read the
DLV Test (D2)
Verify Register
again, after 0.1 seconds, to verify that BIT Testing is activated. This test continuously sequences
between the channels on the card with each output being measured for approximately 180mSec. If the measured
position has an error greater the 0.05º, a flag will be set in the appropriate register. If the input position is stepped
more than 0.05º during a test cycle, the test cycle will not generally indicate an error.
In addition, each DLV Excitation input and signal output is continually monitored. Any failure triggers an Interrupt
(if enabled) and the results are available in the
DLV Signal
and
DLV Excitation Status Registers.
The off-line (D3) Test -
Writing “1” to the D3 bit of the
DLV Test Enable Register
initiates a BIT Test that
generates and tests 72 different positions to an accuracy of 0.05
. External excitation is required and outputs
must be ON. The DLV Status bits will be set to indicate an accuracy problem. Results are available in the
DLV
Test Status Registers
and if enabled, an interrupt will be generated (See
Interrupt Register
). Test cycle takes
about 30 seconds and the D3 bit changes from “1” to “0” when test is complete. The testing requires no external
programming,
and can be terminated at any time by writing a “0” to the D3 bit of the
DLV Test Enable Register.
CAUTION:
Outputs must be ON and Excitation must be supplied during this test. Output is therefore
active. Check connected loads for possible interaction.
D15
D14
D13
D12
D11
D10
D9
D8
D7
D6
D5
D4
D3
D2
D1
D0
Test Enable
X
X
X
X
X
X
X
X
X
X
X
X
D3
D2
X
X
Test (D2) Verify
Card will write 55h at
Test (D2) Verification
register when (D2) is enabled, approximately every one second. User
can clear to 00h and then read again, after approximately one second, to verify that background bit testing is
activated.
DLV Output Mode
The DLV Output Mode register is utilized for selecting either ratio-metric or absolute (fixed) mode voltages. Ratio-
metric Mode, when selected, will cause the output signal voltage of the channel to vary with the input Excitation.
Fixed Mode, when selected, will set the output to a required magnitude that will not vary with excitation input and
set in the DLV Signal Voltage register regardless of the actual input excitation voltage applied. Set register to “0”
for Ratio-
metric Mode. Set register to “1” for Fixed Mode.
DLV 2-Wire or 3/4-Wire Select
Where a
pplicable, write an “01” or “10” (4-Wire = 01; 2-Wire = 10) to each corresponding channel bit pair,
representing a channel commanded output format, of the
DLV 2-Wire or 3/4 Wire Select Register.
D15
D14
D13
D12
D11
D10
D9
D8
D7
D6
D5
D4
CH2
D2
CH1
D0
Synchro / Resolver Select
X
X
X
X
X
X
X
X
X
X
X
X
D3
D2
D1
D0