84
7.5
LEVELING CONNECTOR CALIBRATION
STEP 12
•
Click the Calibration Tab and choose its
Leveling Connector subtab.
STEP 13
•
Adjust the focus knob to maximize the
fringe pattern (i.e. greatest contrast
between black and white fringes).
•
Press the “Step <>” button.
•
Make sure not to cause any vibration
during the scanning process.
STEP 14
•
As instructed on the Leveling Connector
Calibration dialog, rotate the connector in
the mount by approximately 45 degrees
between each of 8 total measurements,
clicking the “Step <>” button after
rotating. Be sure to keep the mount
locked (or relock after rotation if unable to
rotate while locked). Also, refocus
between steps if needed.
•
Make sure not to cause any vibration
during the scanning process.