70
7.2
SCAN OPTIONS TAB
•
Pass/Fail Criteria
: This section contains the current list of scan Criteria for selection
along with a display of the settings in the chosen criteria. At the completion of a
successful scan, the quantitative results are compared against the Pass/Fail Criteria
selected. The list of criteria available are either factory-supplied (industry-standard)
choices or else User-created criteria sets. Both types can be viewed using the
Criteria Add/Edit dialog option of the Tools menu, but only the User-created sets can
be changed or deleted in that dialog.
•
The Pass/Fail criteria for Fiber Undercut can be a fixed number or alternatively a
calculated limit based on measurement results for other geometry parameters
such as Radius of Curvature and Offset, using a predefined equation. If, for a
chosen Criteria, it is determined by equation, there will be a name for that
displayed in this section rather than a number. Clicking on that name will display
the Pass/Fail values in table and graphical form (see images below). For more
information and explanation of equations
TELCORDIA
IEC-2.50
IEC-1.25
Summary of Contents for Connect-Chek CC6000
Page 2: ......
Page 3: ......
Page 4: ......
Page 13: ......
Page 14: ......
Page 16: ......
Page 28: ...C H A P T E R 2 28 SYSTEM PREPARATION...
Page 40: ...C H A P T E R 3 40 NCC PROGRAM...
Page 42: ...42 3 1...
Page 43: ...C H A P T E R 4 43 MENU BAR FUNCTIONS...
Page 45: ...45 4 1...
Page 55: ...55 4 4...
Page 56: ...56 C H A P T E R 5 TOOLBAR FUNCTIONS...
Page 64: ...64 C H A P T E R 6 IMAGE TABS...
Page 67: ...C H A P T E R 7 67 CONTROL TABS...
Page 75: ...75 7 2 SCAN OPTIONS TAB...
Page 82: ...82 7 4 REFERENCE TOOL CALIBRATION...
Page 92: ...C H A P T E R 9 92 APPENDIX...
Page 104: ...104 9 3 SHIPPING INSTRUCTIONS...