11
•
The Report Generate dialog now shows the correct dates when activated, even if
NCC has remained running more than 24 hours.
•
NCC no longer crashes if attempt made to delete first entry in “Add New
Reference” dialog when there are more than one entry in list.
•
In “Add new reference” dialog (under Calibration tab), the CHANGE function now
works as intended.
•
Live image now operates correctly for all types of Windows user accounts.
Previously, the Live image appeared black for users running NCC in a non-
Windows user account lacking Administrator-level rights, only corrected by
launching the Video Control dialog.
•
Date controls in the Report Generation dialog now work correctly for any
operating system Regional Settings language/locality choice.
Changes for NCC v2.51:
•
Redesigned data storage user interface and operation. Includes the following
changes:
a
Addition of a checkbox allowing Operator to activate a data-save
prompt at scan completion.
b
Moved “AutoSave” On/Off controls from Measure tab to Setup tab on
the user interface and renamed it as “Export to Text file”.
c
User can now define location and name of program database file (used
as data source for Scan Reports).
•
Feature added allowing Operator to delete all data records in the program
database (“Clear Data” option in Report menu list).
•
Significantly reduced volatile (System RAM) memory use by program as well as
the amount of hard disk memory required to store each scan’s data into the
program database.
•
Fiber Diameter entry (by Operator) now limited by program to 180 microns
maximum.
•
Changes made to Scan Report format to insure result values and their
corresponding Pass-Fail limits are always shown with a proper number of decimal
digits.
Changes for NCC v2.50:
•
Added Report Configuration and Generation capability and added option for
selecting automatic report generation at scan completion.
•
Company Logos imported into the Report configuration now retain their aspect
ratio (i.e.; are not distorted from original shape) when a report is generated.
•
Incorporated data storage into Microsoft Access database (used as data source
for generated Reports).
•
Entering a Connector ID having both text and numeric content is now allowed.
•
Any Reference Tools issued before version 2.00 release that does not
have the temporary label affixed, must be remarked by reversing the
sign of both X and Y markings in order for it to be used with this
version. For example, if original marking was “X = -0.004, Y =
+0.026”, the marking should be changed to read “X = +0.004, Y =
Summary of Contents for Connect-Chek CC6000
Page 2: ......
Page 3: ......
Page 4: ......
Page 13: ......
Page 14: ......
Page 16: ......
Page 28: ...C H A P T E R 2 28 SYSTEM PREPARATION...
Page 40: ...C H A P T E R 3 40 NCC PROGRAM...
Page 42: ...42 3 1...
Page 43: ...C H A P T E R 4 43 MENU BAR FUNCTIONS...
Page 45: ...45 4 1...
Page 55: ...55 4 4...
Page 56: ...56 C H A P T E R 5 TOOLBAR FUNCTIONS...
Page 64: ...64 C H A P T E R 6 IMAGE TABS...
Page 67: ...C H A P T E R 7 67 CONTROL TABS...
Page 75: ...75 7 2 SCAN OPTIONS TAB...
Page 82: ...82 7 4 REFERENCE TOOL CALIBRATION...
Page 92: ...C H A P T E R 9 92 APPENDIX...
Page 104: ...104 9 3 SHIPPING INSTRUCTIONS...