REFERENCE TOOL CALIBRATION
STEP 4
•
Click the Calibration Tab and choose its
Reference Tools subtab.
STEP 5
•
Select a previously saved Reference Tool
ID from the drop down menu. When a
reference tool choice is selected, its
corresponding Offset X and Y values will
appear beneath the selection box. if
there are no X and Y factors matching
those printed on the Reference Tool, a
new Reference Tool ID must be created
for it by the production supervisor via the
Reference Tool Add/Edit function in the
program’s Tools menu.
STEP 6
•
Adjust the focus knob to maximize the
fringe pattern (i.e. greatest contrast
between black and white fringes).
•
Press the Calibrate button.
•
Make sure not to cause any vibration
during the scanning process.