Setting Parameters
page 123
NF 650 - rev. 01 / 29.06.2020
34.5.3 Feature unlock
Use this function to unlock (optionally available) features provided that they have been ordered
from NIVUS.
Â
Ü
Feature unlocking procedure:
1. Click button >Feature Unlock<.
2. In the menu opening up click >Feature Unlock< button.
3. Enter function code and confirm with "Input".
The transmitter confirms the feature unlock indicating "Successful". The assigned
licence is shown in the display.
4. The unit requires a restart. Subsequently, the new features are available in the
according menus and now can be adjusted and used.
Fig. 34-10 Feature unlock
34.5.4 Reboot
A transmitter reboot will interrupt the current measurement process.
The system will reboot using the parameters set and saved last. After booting, the system
behaves as when the instrument is switched on (like a PC).
This menu point replaces the system shutdown and reboot.
All parameters, counters and saved data are retained.
Fig. 34-11 Reboot
34.5.5 Restart Measurement
This option interrupts the currently running measurement process and starts a new measure-
ment.