15
PREPARATION
1. Diopter Adjustment
( 1) Push the mirror block changeover lever to put the
brightfield mirror block (with B.F. indication) in
the optical path.
(2) Push both the OB/PHOTO•SIDE and OB/PHOTO
changeover levers to provide the binocular
eyepieces with 100% light output.
(3) Tum on the power switch of the power supply
unit. Rotate its lamp intensity control dial to
CTRL. Set the lamp intensity control dial of the
microscope base at about 6V.
(4) If a Nomarski prism, the polarizer and the
analyzer are in the optical path, remove them
from the optical path.
(5) Fully open the aperture and field diaphragms.
Fully open
the aperture
diaphragm.
Fully open the
field diaphragm.