IV. Assembly
57
5. Filter slider
Attach the filters to the filter sliders. (Push aside the claw on the
circular mount when attaching the filter.)
Insert the sliders on the dia-illuminator.
6. Stage
Place the stage on the mounts at the center front of the base and the dia-illuminator.
(If your stage is equipped with the movement handle, let it position at the right rear of the microscope though it can also
be positioned at front left.)
Fix the stage with four M5 hexagonal bolt supplied together with the stage.
7. Lamp installation (lamp exchange)
•
The combination of the lamp, dia-illuminator and the power supply is fixed. See page 38 and use them in
correct combination. Be sure to use the specified lamp.
•
When exchanging the lamp, turn off the power and unplug the power cord to avoid electric shock.
•
The lamp and its surroundings become extremely hot during lamp illumination and immediately after the
lamp is distinguished. When exchanging the lamp, turn off the lamp and wait till the lamp cools
sufficiently.
•
Do not touch the lamp with bare hands. Dusts and fingerprints on the lamp surface will lead to uneven
illumination and shortening of the lamp life. Use gloves.
When using “T-DH dia-illuminator 100
W
W
W
W
”:
Remove the cover of the lamphouse by loosening the clamp
screw on the right side of the cover.
Securely insert the lamp to the lamp socket till the limit while
pressing down the lamp clamp lever.
Slowly release the lamp clamp lever. Make sure that the lamp
is not tilted. If so, reattach the lamp.
Return the cover to the lamphouse and fix it by the clamp
screw.
Cover clamp screw
Lamp clamp lever
Caution for heat:
Do not touch while the lamp
is lit or immediately after the
lamp is distinguished
When using “T-DS dia-illuminator 30
W
W
W
W
”:
Open up the cover at the rear part of the dia-illuminator.
Attach the lamp to the socket.
Close the cover
Caution for heat:
Do not touch
while the lamp is
lit or immediately
after the lamp is
distinguished