74
Required tools
•
Two hexagonal screwdrivers (2 mm) (provided with the microscope)
•
Hexagonal wrench (3 mm) (provided with the microscope)
When these tools are not used, place them in the tool holder at the right side of the microscope base.
Installation location
This product is a precision optical instrument. So, the usage or storage in an inappropriate
environment may result in malfunctions or poor performance. Consider the following factors when
selecting an installation location:
•
Avoid a brightly lit location, such as exposed to direct sunlight or directly under a room light.
If there is excessive ambient light, the image quality deteriorates.
•
Always install the microscope with a surrounding clear area of 10 cm or more.
•
Install the product in a location that is free from considerable dust or dirt.
•
Install the product on a flat surface with little vibration.
•
Install the product on a sturdy desk or table that is able to bear the weight of the instrument.
•
Do not install the product in a hot and humid location.
•
Arrange a layout that allows easy removal of the power cord from the inlet of the product in
the event of an emergency.
•
For details about the operating environment and storage environment, see “VII.
Specifications.”
Combination of the illuminator and the light source
This product is UL-listed only in the combination of the illuminator and the light source (the lamp
house and the lamp) describe below. Please take note that if a light source other than the specified
ones are installed onto this microscope, this microscope system will not be treated as a UL-listed
product.
•
Illuminator:
LV-UEPI2A Motorized Universal Epi Illuminator 2A
•
Lamp house:
LV-LH50PC Precentered Lamp House
(both for the episcopic illumination and for the diascopic illumination)
•
Lamp:
LV-HL50W 12V 50W LONGLIFE halogen lamp, or non-Nikon 12V 50W
SHORTLIFE halogen lamp (model name: OSRAM HLX 64610, OSRAM
HLX 64611, or PHILIPS 7027).