90
2. Connecting the RS-232C cable
Connect the RS-232C cable to the product and to the external light source (Intensilight C-HGFIE or
EXFO X-Cite 120 PC). With this connection, you can control the shutter and the brightness on the
light source by operating the switches on the microscope.
1
Check that the power supplies for the microscope and
the light source are turned off.
2
Connect the RS-232C cable provided with the light
source to the RS-232C connector on the product and
to the RS-232S port on the light source.
Screws are provided for these connectors. Make sure
to fix these connectors with screws.
JAP
AN
UEPI2A
USB
RS232C
LCNT
ND8
NCB
F.S.
Achr N.A = 0.9
JAPAN
0.8
0.7 0.6
0.5 0.4 0.3 0.2 0.1
3 x 2 S
TAGE
JAPAN
F
.
STOP
JAPAN
BF
DF FL1 FL2
FL1
FL2
100
20
0
100
IN
OUT
LV-TT
2
INTENSILIGHT
C-HGFIE
LAMP
POWER
RUN
TIME hrs
.
Power cord
RS-232C cable
Light guide fiber
External light source
(Intensilight C-HGFIE)
Optical fiber adapter
T h i s d ev i c e c o m p l i e s w i t h Pa r t 1 5
o f t h e F C C R u l e s . O p e r a t i o n i s s u b j e c t
t o t h e fo l l o w i n g t w o c o n d i t i o n s :
( 1 ) T h i s d ev i c e m ay n o t c a u s e h a r m f u l
i n t e r fe r e n c e , a n d ( 2 ) t h i s d ev i c e m u s t
a c c e p t a n y i n t e r fe r e n c e r e c e i ve d ,
i n c l u d i n g i n t e r fe r e n c e t h a t m ay
c a u s e u n d e s i r e d o p e r a t i o n .
T h i s C l a s s A d i g i t a l a p p a r a t u s
c o m p l i e s w i t h C a n a d i a n I C E S - 0 0 3 .
C e t a p p a r e i l n u m é r i q u e d e l a
c l a s s e A e s t c o n f i r m e à l a n o r m e
N M B - 0 0 3 d u C a n a d a .
JAPAN
C-HGFIE
8 7 0 0 0 1
4N75
INSPECTION
EQUIPMENT
geprüfte
Sicherheit
TÜV
Rheinland
Product Safety
AC-IN 100-240~
50/60Hz 2.0A
RS-232C
REMOTE
SHUTTER
Light guide
port
RS-232C
port
Rear view of the Intensilight
C-HGFIE
Operation for the light source
For details about the operation for the external light source, refer to the instruction manual
for the Intensilight C-HGFIE or the EXFO X-Cite 120 PC.
When the external light source is connected with the product, the shutter and the brightness
(iris) on the light source can be controlled only by operating the switches on the operation
panel of the microscope. For the EXFO X-Cite 120 PC, if a button on the light source is
touched, “LOC”c is displayed on the LCD display for two seconds indicating that the
operation on the light source is disabled.