Lambda Plate Slider for the Diascopic Illumination
The sensitive color polarization microscopy under
the diascopic illumination can be performed when
the lambda plate slider is inserted into the slot on
the nosepiece and the lambda plate is placed into
the optical path in the configuration of the
polarization microscopy under the diascopic
illumination. The sensitive color DIC microscopy
under the diascopic illumination can be performed
when the lambda plate slider is inserted into the slot
on the nosepiece and the lambda plate is placed into
the optical path in the configuration of the DIC
microscopy under the diascopic illumination.
Attaching/removing the lambda plate
Attach the lambda plate to the slot for the lambda plate
slider of the LV-NU5AI. (see Page 80.)
Loosen the slider limiter screw with the hexagonal
screwdriver provided with the microscope and insert the
lambda plate slider into the slot. And then, tighten the
slider limiter screw until it is inserted into the groove.
To remove the lambda plate slider from the nosepiece,
loosen the slider limiter screw sufficiently with the
hexagonal screwdriver and pull out the llambda plate
slider.
Placing the lambda plate into the optical path
When the lambda plate slider is pushed into the second click-stop position, the lambda plate is
placed into the optical path.
Removing the lambda plate from the optical path
When you wish to remove the lambda plate from the optical path, pull the lambda plate slider to the
first click-stop position.