Bright-field Microscopy under the Dia Illumination
1
Turn on the power switch.
When the power to the microscope is turned on, the
microscope starts initialization. And then, the power
indicator on the microscope base is lit. (See Page 34.)
2
Set the microscope for the bright-field microscopy under the diascopic illumination.
If accessories (*1 to *4) for the DIC microscopy under the episcopic illumination are in place, remove
them from the optical path.
1
Push in the optical path selector lever and make
the distribution of light for the binocular part
100%. (See Page 48.)
2
Press the CUBE switch on the operation panel
and light up the “DF (dark-field)” position of the
microscopy method indicator. (See Page 37.)
To use the diascopic illumination, be sure to set
the LV-UEPI2A to “DF.”
3
Press the OBJ. switch on the operation panel and
locate the 10x objective into the optical path.
(See Page 43.)
4
Press the DIA switch on the operation panel to
light up the diascopic illumination lamp, and
adjust the brightness with the DIA brightness
switch. (See Page 42.)
5
Locate the NCB11 filter into the optical path and
compensate the color temperature. (See Page 44.)
6
Adjust the brightness with ND filters.
(See Page 44.)
7
Rotate the field diaphragm control toward you
and fully open the field diaphragm for the
diascopic illumination. (See Page 54.)
8
To use the universal condenser, set the condenser
turret to the “O” position. (See Page 55.)
9
Rotate the condenser focus knob to focus the
condenser. (See Page 53.)
10
Rotate the condenser aperture diaphragm ring
toward the left to fully open the condenser
aperture diaphragm. (See Page 54.)
UEPI2
A
USB
RS23
2C
LCNT
ND8
NCB
F.S.
Achr N.A = 0.9
JAPAN
0.8
0.7 0.6
0.5 0.4 0.3 0.2 0.1
3 x 2 S
TAGE
JAPAN
F
.
STOP
JAPA
N
BF
DF FL1 FL2
FL1
FL2
100
20
0
100
IN
OUT
LV-TT
2
OBJ.
CUBE
A.S.
EPI
DIA
EPI
DIA
5
7
6
4
2
4
3
2
3
1
2
Set the microscope for the diascopic bright-field
microscopy.
*
3
*
4
*
1
*
2
DIA brightness switch
CUBE switch
DIA switch
OBJ. switch
Operation
panel
OBJ.
CUBE
A.S.
EPI
DIA
EPI
DIA
Achr N
.A
0.9
0.2
0.3
0.4
0.5
0.6
0.7
0.8
=
3 x 2
STA
GE
JAPA
N
Power
indicator
Turn on
the power.
1
OBJ.
CUBE
A.S.
EPI
DIA
EPI
DIA
3 x 2
STA
GE
JAPA
N
9
8
10
OBJ.
CUBE
A.S.
EPI
DIA
EPI
DIA
Achr N
.A
0.9
0.2
0.3
0.4
0.5
0.6
0.7
0.8
=
3 x 2
STA
GE
JAPA
N
9
10
Universal condenser
Other condenser