HOW TO USE - Preparation

Confirm that the Contact Point and Rear Cover are tightly fastened. 

If loose, tighten Contact Point and Rear Plate Screws.

Attach to the Gauge Holder using the Stem or Rear Lug. 

Mounting Gauge by other than Stem or Lug will cause inaccuracy and product damage.

Confirm that Pointer and Revolution Counter movement is smooth. 

Using fingertip, gently press on Contact Point to move the Spindle up and down. Motion of Pointers should be 
smooth. If it is not smooth make sure Stem is not clamped too tight, and adjust. Also make sure Pointer is stable 
at set position.

Make sure Spindle axis is perpendicular to measured surface. 

If Spindle (Contact Point) is not at a right angle to 
surface, Gauge will not operate properly and 
measurement will be inaccurate. Always keep the 
Spindle axis perpendicular.

 ※

When used to check parallelism of Milling 
Machine vise, use a Magnetic Base to mount the 
Gauge perpendicular to surface, and move it out 
of the way during operation to prevent 
interference.

HOW TO USE - Comparison Measurements

Set Up Reference Part.

Carefully lift Spindle with fingertip, and, taking care not to 
hit Spindle from the side, insert the Reference Part or 
Master under Contact Point.

Set the Origin.

Adjust the gauge mount or rotate the Bezel to set the 
Gauge to "0".

Remove Reference, and begin measurements. 

Remove Reference or Master, careful not to shock Spindle. 
Insert part to be measured and read the measurement off 
the Scale. 

 ※

Setting the Limit Markers

        Limit Markers can be moved to show acceptance range 

for measurements.

HOW TO USE - Parallelism, Flatness, Runout, etc.

Position Contact Point on surface. 

Carefully lifting Spindle with fingertip, and taking care not to hit Spindle from the side, position the surface to be 
measured under the Contact Point.

Set the Origin. 

Adjust the gauge mount, or rotate the Bezel to set the Gauge to "0".

Read the scale as the measured part is moved. 

Slowly move the part while monitoring the Pointer and reading the measurement.

Right Angle

Pointer

Limit Markers

Ex.) Pass Range : 
       

0.05 to 

0.15mm

Measured Object

Measured Object

Summary of Contents for PROFESSIONAL DI-10KD

Page 1: ...n the following page Limit Markers For marking acceptance range for Pass Fail testing Pointer Needle Shows measurement on Main Scale Main Scale Main Scale rotate Bezel to turn Bezel Clamp Screw Clamps Scale to prevent rotation Revolution Counter Pointer indicates Main Scale 100 Rev Counter Scale Each division is Main Scale 100 Stem For holding and mounting Gauge Spindle Shaft moves up and down wit...

Page 2: ...n and unauthorized people Use in location contrary to the above may cause poor accuracy damage to the product or may result in accident or injury Handle With Care Do not drop or subject to shock do not place under heavy objects Damage may cause failure or poor accuracy Make sure Gauge holder is rigid Holder must be sufficiently secure to prevent deflection from the weight of the Gauge Holder must ...

Page 3: ...nd move it out of the way during operation to prevent interference HOW TO USE Comparison Measurements Set Up Reference Part Carefully lift Spindle with fingertip and taking care not to hit Spindle from the side insert the Reference Part or Master under Contact Point Set the Origin Adjust the gauge mount or rotate the Bezel to set the Gauge to 0 Remove Reference and begin measurements Remove Refere...

Page 4: ...there is no force on the Spindle such as pushed in or lateral force Keep away from moisture and direct sunlight and secure from unauthorized personnel I360 K 1803 Niigata seiki Co Ltd 6 15 22 Tsukanome Sanjo Niigata Japan 955 0055 Tel 81 256 31 5670 Fax 81 256 39 7730 MAIL intl sales niigataseiki co jp URL http www niigataseiki co jp CALIBRATION Replacement Contact Point Part No DI 10KD DI 10 DI C...

Reviews: