TROUBLESHOOTING

SPECIFICATIONS

Origin position shifts during measurement.

Temperature changes during measurement can cause repeatability error. Please try the following solutions:

   

Use in location with constant temperature.

   

When taking measurements, periodically adjust zero point using a Master reference to correct for temperature 

induced drift.

Measurement is not stable, or measurement accuracy is poor.

Contact Point may be worn.  

Worn Contact Point will affect accuracy. Periodically check for wear, and replace if wear is affecting measurement 
accuracy.

Remove any dust or dirt after use.  

Do not lubricate.

Wipe any contamination from Spindle sliding surface using a dry cloth, or cloth moistened with alcohol.

To clean other surfaces, wipe with a soft dry cloth, or a cloth moistened with a mild cleaner.

Check for wear of Contact Point.

Measurement accuracy will be affected by worn Contact Point. Regularly check for wear and replace 
Contact Point if worn.

Store in provided case in a cool, dark, and dry location.

During storage, make sure there is no force on the Spindle (such as pushed in, or lateral force.)

Keep away from moisture and direct sunlight, and secure from unauthorized personnel.

I360-K 1803

Niigata seiki Co., Ltd.

6-15-22, Tsukanome, Sanjo, Niigata, Japan, 955-0055
Tel. : +81-256-31-5670   Fax. : +81-256-39-7730

MAIL   [email protected]

URL    http://www.niigataseiki.co.jp

CALIBRATION

Replacement Contact Point Part No.

DI-10KD

DI-10

DI-CPK

DI-1058

DI-CP

  

        

                                  WDI-0158SP

WDI-0558SP

WDI-1058SP

WDI-W

AFTER USE CARE, STORAGE

To maintain measurement accuracy, periodic calibration is recommended.

(For reference, we recommend a calibration interval of 3~4 months when used in a factory.)

Outside Japan, Please contact distributor or place of 
purchase to inquire about calibrations services.

DI-10KD
DI-10
DI-1058
WDI-0558SP
WDI-1058SP
WDI-0158SP

0.01
0.01
0.01
0.01
0.01
0.001

5
5
5
3
3
2

1.5N

1.5N

1.5N

1.5N

1.5N

2N

1
1
1
1
1
0.2

0

10

0

10

0

10

0

  5

0

10

0

  1

±

0-100

±

0-100

±

0-100

±

0-100

±

0-100

0-100-0

Model No.

Scale

Contact

Point

125
125
180
190
190
190

Weight

(g)

Graduation

(mm)

Measurement

Range

(mm)

Retrace

Error

(

μ

m)

Measuring 

Force

(N)

Range

(1 rev.)

(mm)

8
8
8
8

10

3

±

20

±

20

±

20

±

16

±

20

±

  5

5
5
5
3
3
0.3

Adjacent

Error

(1/10rev.):

(

μ

m)

Full Range

Accuracy

(

μ

m)

Repeat 

Accuracy

(

μ

m)

Carbide

Ball

Steel 

Ball

Summary of Contents for PROFESSIONAL DI-10KD

Page 1: ...n the following page Limit Markers For marking acceptance range for Pass Fail testing Pointer Needle Shows measurement on Main Scale Main Scale Main Scale rotate Bezel to turn Bezel Clamp Screw Clamps Scale to prevent rotation Revolution Counter Pointer indicates Main Scale 100 Rev Counter Scale Each division is Main Scale 100 Stem For holding and mounting Gauge Spindle Shaft moves up and down wit...

Page 2: ...n and unauthorized people Use in location contrary to the above may cause poor accuracy damage to the product or may result in accident or injury Handle With Care Do not drop or subject to shock do not place under heavy objects Damage may cause failure or poor accuracy Make sure Gauge holder is rigid Holder must be sufficiently secure to prevent deflection from the weight of the Gauge Holder must ...

Page 3: ...nd move it out of the way during operation to prevent interference HOW TO USE Comparison Measurements Set Up Reference Part Carefully lift Spindle with fingertip and taking care not to hit Spindle from the side insert the Reference Part or Master under Contact Point Set the Origin Adjust the gauge mount or rotate the Bezel to set the Gauge to 0 Remove Reference and begin measurements Remove Refere...

Page 4: ...there is no force on the Spindle such as pushed in or lateral force Keep away from moisture and direct sunlight and secure from unauthorized personnel I360 K 1803 Niigata seiki Co Ltd 6 15 22 Tsukanome Sanjo Niigata Japan 955 0055 Tel 81 256 31 5670 Fax 81 256 39 7730 MAIL intl sales niigataseiki co jp URL http www niigataseiki co jp CALIBRATION Replacement Contact Point Part No DI 10KD DI 10 DI C...

Reviews: