Page 2: ...Ku Tokyo 113 0033 Japan Telephone 81 3 5844 2641 Facsimile 81 3 5844 2642 NIDEK INCORPORATED 47651 Westinghouse Drive Fremont California 94539 U S A United States Agent Telephone 1 510 226 5700 Facsi...

Page 3: ...of Data Output 24 3 COMMUNICATION PROCEDURE other than NCP20 51 3 1 Outline 51 3 2 Communication Control Signals DTR Output to DSR Input 52 3 3 Actual Communications 55 3 3 1 The LM requests the compu...

Page 4: ...am Example 87 6 3 Confirmation and Setting of Receiver Computer 88 6 3 1 Reception structure of measurement data 88 6 3 2 Setting up the shared folder 88 6 3 3 Confirming items necessary for other set...

Page 5: ...rding to the communica tion method RS 232C interface RS 232C compliance USB interface USB 2 0 HOST FUNC compliance LAN interface 10 100BASE T Ethernet Priority of interface RS 232C USB The communicati...

Page 6: ...2 OUTLINE OF COMMUNICATION Outline of Communication Function MLM26 RTZ002F...

Page 7: ...set the settings are unchangeable The settings of Parameter 5 Baud rate Parameter 6 Parity and Parameter 8 Stop bits are changeable Be sure to set the Auto print S and Auto print R L parameters to Of...

Page 8: ...0 to 255 factory setting 0 3 Request 0 1 to 10 0 seconds factory setting 2 0 4 Reception 1 to 100 seconds factory setting 10 5 Response 0 1 to 10 0 seconds factory setting 2 0 6 Re request 0 1 to 10...

Page 9: ...Off On Display Off The measured prism value is not transmitted at any time On The measured prism value is transmitted at all times When the prism display is set to BU D BI O setting the prism value i...

Page 10: ...anual provided with the computer Use a special interface cable that is available as an option OPIF 11 5 m between the DIN 8 pin and D Sub 9 pin connectors if necessary 1 2 3 4 5 6 7 8 1 9 5 6 Signal n...

Page 11: ...re not performed This section explains the formats of these commands or measurement data Symbols other than ordinary characters in commands or data strings are described as com munication control char...

Page 12: ...nds this command to the LM when the computer is ready for the reception of mea surement data 3 RS command LM Computer Command for the LM to request the computer to receive the measurement data and to...

Page 13: ...number specified by the AB command is in effect until it is reset on the LM nn Abbe number 20 to 60 e g 34H 31H for number 41 6 CH command Computer LM Command for the computer to request the LM to spe...

Page 14: ...request the UV measurement and send the lens power When the UV parameter is set to Off UV measurement is disabled When it is set to With only the reading of the lens power is performed 11 IZ command...

Page 15: ...adecimal numbers in the low order two bytes are represented by four byte ASCII codes as a simple sum of SOH 01H at the header of the data to EOT 04H Be aware that CR code 0DH is excluded from the calc...

Page 16: ...is output When both left and right eye lenses are measured the data is output in following order Right Left In the order of 1 4 Right 5 15 data Left 5 15 data Measured lens mode Prism layout Lens 4 Si...

Page 17: ...umber data is output when RT data No is valid or when RT data No is invalid and Print number parameter is On Number 0001 9999 The print number is counted up every time it is cleared when the Print num...

Page 18: ...1 50 D AXIS 180 Identification code Single Space space Right R Space R Left L Space L SPH Four digits with a fixed decimal point and with a polarity sign CYL Four digits with a fixed decimal point an...

Page 19: ...ith no polarity sign and the leading digit 0 remaining When the Near parameter is set to Near SPH near SPH data is added to the end of the addition powers Both the addition power and near SPH data are...

Page 20: ...is set to BU D BI O The measurement data is output as fixed four digits including a decimal point with no polarity sign and the leading digit 0 remaining In addition a character representing the base...

Page 21: ...as two digits with the leading digit 0 remaining 1 Example of single data V Length 12 mm Identification code Single P B P Space B Space Right PR BR Left PL BL PRISM Four digits with a fixed decimal po...

Page 22: ...ement position 8 mm from the add power start point to near portion 11 Green transmittance data The measurement data of green transmittance is output The measurement data of green transmittance is outp...

Page 23: ...e is output as three digits with the leading digit 0 remaining 1 Example of the single data UV 20 2 Example of left data UV 15 G 0 8 0 ETB CR 47H 20H 30H 38H 30H 17H 0DH Green 3 characters G R 0 7 5 E...

Page 24: ...asurement data of inside amount is output Inside amount is only output when the distance portion of Right or Left PD is measured Unmeasured inside amount is output as The data is sent in the order of...

Page 25: ...f BI O BU D setting as fixed four digits including a decimal point with the leading digit 0 remaining In addition as with the Prism parameter a character representing the base direction is added to th...

Page 26: ...success when the initialization is successful Since this format is the same as that of data the success can be processed with normal interface soft ware The data identification code shows the detail o...

Page 27: ...s of the operator s manual In the case of measurement error SOH D L M STX I E ETB CR EOT CR 01H 44H 4CH 4DH 02H 49H 45H 17H 0DH 04H 0DH The CR codes are added Character representing Error Character re...

Page 28: ...ent conditions Single vision lens measurement with the single state designated SPH 1 00D CYL 0 00D AXIS 0 No prism measurement performed No transmittance measurement performed CR code None Operator pa...

Page 29: ...d 01H 44H 4CH 4DH 02H I D N I D E K L M 1 8 0 0 P d ETB To be continued 49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 38H 30H 30H 50H 64H 17H Company name Model name LM 1800PD D A 2 0 1 0 1 1 2 3 1...

Page 30: ...ed to the end of ETB and EOT SOH D L M STX To be continued 01H 44H 4CH 4DH 02H I D N I D E K L M 1 8 0 0 P d ETB CR To be continued 49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 38H 30H 30H 50H 64H...

Page 31: ...be continued 01H 44H 4CH 4DH 02H I D N I D E K L M 1 8 0 0 P d ETB CR To be continued 49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 38H 30H 30H 50H 64H 17H 0DH Company name Model name LM 1800PD D A...

Page 32: ...OH D L M STX To be continued 01H 44H 4CH 4DH 02H I D N I D E K L M 1 8 0 0 P d ETB CR To be continued 49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 38H 30H 30H 50H 64H 17H 0DH Company name Model nam...

Page 33: ...left eye lens data is output after the transmission of the right eye lens data SOH D L M STX To be continued 01H 44H 4CH 4DH 02H I D N I D E K L M 1 8 0 0 P d ETB CR To be continued 49H 44H 4EH 49H 4...

Page 34: ...2DH 30H 32H 2EH 30H 30H 2DH 30H 30H 2EH 35H 30H 31H 38H 30H 17H 0DH Left SPH Left CYL Left AXIS A L 0 2 2 5 ETB CR P L 0 1 2 5 O ETB CR To be continued 41H 4CH 30H 32H 2EH 32H 35H 17H 0DH 50H 4CH 30H...

Page 35: ...de Added This is an example of the LM 1800P of the maximum data size 233 characters in normal lens measurement mode When the Near parameter is set to Near Sph both the addition powers ADD and near sph...

Page 36: ...35 N R 0 0 7 5 ETB CR 0 1 7 5 ETB CR To be continued 4EH 52H 2BH 30H 30H 2EH 37H 35H 17H 0DH 2BH 30H 31H 2EH 37H 35H 17H 0DH Right NSPH Right NSPH2 140 145 150 155 P R 0 2 2 5 I ETB CR P R 0 2 0 0 D E...

Page 37: ...H 2EH 35H 30H 17H 0DH Left NSPH Left NSPH2 225 230 235 240 P L 0 1 2 5 O ETB CR P L 0 2 0 0 U ETB CR To be continued 50H 4CH 30H 31H 2EH 32H 35H 4FH 17H 0DH 50H 4CH 30H 32H 2EH 30H 30H 55H 17H 0DH Lef...

Page 38: ...imum data size 233 characters in progressive power lens measurement mode The checksum is added according to the setting of the Com mode parameter NCP10 When the Near parameter is set to Near Sph both...

Page 39: ...17H 0DH 50H 52H 30H 32H 2EH 30H 30H 44H 17H 0DH Right PRISM1 Right PRISM2 145 150 155 G R 0 8 0 ETB CR U R 0 2 0 ETB CR To be continued 47H 52H 30H 38H 30H 17H 0DH 55H 52H 30H 32H 30H 17H 0DH Right Gr...

Page 40: ...r NCP10 The SE value is output The addition power is not measured When reading two kinds of barcode data Operator Patient ID 1 5 Characters SOH D L M STX To be continued 01H 44H 4CH 4DH 02H 10 15 20 I...

Page 41: ...H 30H 32H 2EH 30H 30H 44H 17H 0DH Right PRISM2 135 140 145 G R 0 8 0 ETB CR U R 0 2 0 ETB CR To be continued 47H 52H 30H 38H 30H 17H 0DH 55H 52H 30H 32H 30H 17H 0DH Right Green Right UV 150 155 160 16...

Page 42: ...P10 Prism value is the measured prism power When reading two kinds of barcode data Operator Patient ID 1 5 Characters SOH D L M STX To be continued 01H 44H 4CH 4DH 02H 10 15 20 I D N I D E K L M 1 8 0...

Page 43: ...G R 0 8 0 ETB CR U R 0 2 0 ETB CR To be continued 47H 52H 30H 38H 30H 17H 0DH 55H 52H 30H 32H 30H 17H 0DH Right Green Right UV 140 145 150 155 L 0 2 0 0 0 0 5 0 1 8 0 ETB CR To be continued 20H 4CH 2D...

Page 44: ...mm Net prism measurement Net H Pri IN 1 25 Net V Pri UP1 00 CR code Added This is an example of LM 1800PD of the maximum data size 300 characters in normal lens measurement mode When the Near paramet...

Page 45: ...ht AXIS 105 110 115 A R 0 2 0 0 ETB CR 0 3 0 0 ETB CR To be continued 41H 52H 30H 32H 2EH 30H 30H 17H 0DH 30H 33H 2EH 30H 30H 17H 0DH Right ADD Right ADD2 120 125 130 135 N R 0 0 7 5 ETB CR 0 1 7 5 ET...

Page 46: ...R To be continued 50H 4CH 30H 31H 2EH 32H 35H 4FH 17H 0DH 50H 4CH 30H 32H 2EH 30H 30H 55H 17H 0DH Left PRISM1 Left PRISM2 245 250 255 G L 0 8 0 ETB CR U L 0 2 0 ETB CR To be continued 47H 4CH 30H 38H...

Page 47: ...PH 0 25 D PRISM OUT 1 25 UP 2 00 BU D BI O setting Transmittance measurement Green 80 UV 20 Progressive length 17 mm Channel width Positions of 10 mm 18 mm LPD 32 5 mm LINS 2 0 mm Total PD 64 0 mm CR...

Page 48: ...90 95 100 R 0 1 2 5 0 0 7 5 1 2 0 ETB CR To be continued 20H 52H 2DH 30H 31H 2EH 32H 35H 2DH 30H 30H 2EH 37H 35H 31H 32H 30H 17H 0DH Right SPH Right CYL Right AXIS 105 110 115 120 A R 0 2 0 0 ETB CR...

Page 49: ...H 50H 4CH 30H 32H 2EH 30H 30H 55H 17H 0DH Left PRISM1 Left PRISM2 230 235 240 D L 1 7 ETB CR W L 1 0 1 8 ETB CR To be continued 44H 4CH 31H 37H 17H 0DH 57H 4CH 31H 30H 2FH 31H 38H 17H 0DH Left progres...

Page 50: ...er and PD are not measured When reading two kinds of barcode data Operator Patient ID 1 5 Characters SOH D L M STX To be continued 01H 44H 4CH 4DH 02H 10 15 20 I D N I D E K L M 1 8 0 0 P d ETB CR To...

Page 51: ...ETB CR U R 0 2 0 ETB CR To be continued 47H 52H 30H 38H 30H 17H 0DH 55H 52H 30H 32H 30H 17H 0DH Right Green Right UV 150 155 160 165 L 0 2 0 0 0 0 5 0 1 8 0 ETB CR To be continued 20H 4CH 2DH 30H 32H...

Page 52: ...ta is the measured prism value When reading two kinds of barcode data Operator Patient ID 1 5 Characters SOH D L M STX To be continued 01H 44H 4CH 4DH 02H 10 15 20 I D N I D E K L M 1 8 0 0 P d ETB CR...

Page 53: ...2H 30H 32H 30H 17H 0DH Right Green Right UV 140 145 150 155 L 0 2 0 0 0 0 5 0 1 8 0 ETB CR To be continued 20H 4CH 2DH 30H 32H 2EH 30H 30H 2DH 30H 30H 2EH 35H 30H 31H 38H 30H 17H 0DH Left SPH Left CYL...

Page 54: ...50 RS 232C Communication and USB Communication Example of Data Output MLM26 RTZ002F...

Page 55: ...the data on the screen or saves it to a floppy disk Therefore the computer needs a program that sends or receives these commands or data and analyzes displays or saves the received data In this manner...

Page 56: ...ice is not ready for communication during printout by the LM or the processing of the previous data by the com puter it sends DTR OFF to the other device to allows the other device to know the status...

Page 57: ...of 0 1 second when Com mode is set to NIDEK or 4 seconds when Com mode is set to NIDEK2 a timeout occurs in the LM and the LM terminates communication The LM analyzes Command from the computer and pr...

Page 58: ...R is enabled ON on the LM If the computer is put into the wait state to wait for Data from the LM State 7 both the com puter and LM are put into the wait sate which appears to the operator that the co...

Page 59: ...Setting the LM Set the parameters related to the RS 232C interface on the LM in accordance with 2 1 Basic Interface Specifications Page 3 to conform them to the settings on the computer 4 Connecting...

Page 60: ...that DSR is enabled DTR of the LM ON and enables DSR ON 3 LM When the LM enables DSR DTR of the computer ON it sends RS command to the computer to request the computer to send SD command 4 COM When th...

Page 61: ...that it is starting communication 3 LM When the LM confirms that DSR DTR of the computer is enabled ON it enables DTR On to wait for Command from the computer 4 COM When the computer conforms that DSR...

Page 62: ...measurement data 6 COM The computer sends SD command to request the LM to send measurement data after waiting for about 0 1 second until the LM saves the measurement data 7 LM After receiving SD comma...

Page 63: ...o On 1Line 1000 to 1290 When the LM requests the computer to receive the measurement data For using the print button or export button to send the measurement data to the computer Procedure 1 Enter the...

Page 64: ...data is displayed on the screen 6 After the transmission of data the measurement data in the LM is cleared Note 1 Lines 1100 2110 and 3130 in the sample program have a command to open the RS 232C port...

Page 65: ...I 1 TO 210 LM I NEXT I buffer clear 1090 1100 OPEN COM O81NN AS 1 RS 232C DTR on 1110 OUT H32 H15 DTR off 1120 PRINT PRINT Let s Measure and LM s PRINT sw on 1130 IF INP H32 AND H80 H0 THEN 1130 check...

Page 66: ...40 CHILD SWITCH RS232C 0 9600 set 9600 bit sec 3050 DIM LM 210 3060 RD CHR 1 CLM CHR 2 RD CHR 23 CHR 4 RD command 3070 SD CHR 1 CLM CHR 2 SD CHR 23 CHR 4 SD command 3080 CL CHR 1 CLM CHR 2 CL CHR 23 C...

Page 67: ...190 1210 1260 1290 END Initialization Opening the RS 232C port Disabling DTR that was enabled by OPEN The LM enables DTR ON and waits for the start of communication The computer enables DTR ON and sta...

Page 68: ...SR ON Initialization Opening the RS 232C port DTR is left en abled ON because communication is started immediately The LM enables DTR ON and waits for the start of communication Transmission of the SD...

Page 69: ...e RS 232C port DTR is left en abled ON because communication is started immediately The LM enables DTR ON and waits until the communication is ready Transmission of the RD command The computer waits f...

Page 70: ...66 COMMUNICATION PROCEDURE other than NCP20 Flowchart MLM26 RTZ002F...

Page 71: ...t numbers are compared and the port with large number is preceded The numbers can be changed 0 to 255 Destination port No This is set when the destination port No is necessary for specific network con...

Page 72: ...ata are sent When single data is sent the command cord is 0x0000 Only single data is used for the lens meter Message length Indicates the message and block length 2 bytes of the present block Message...

Page 73: ...sends Accept to A to inform the receiving of 1 block of data 5 Repeat the steps from 3 to 4 until all the data is sent When equipment A sends data to B by request from equipment B 1 Equipment B reque...

Page 74: ...hat there are no opponent instruments T O range 0 1 to 10 seconds Initial setting 2 seconds 2 Time out for receiving data Receive T O Set the time out period for receiving all the data or character in...

Page 75: ...002F 4 When equipment A sends data to B by request from equipment A Request to Send Accept to Send Accept Data Data Accept Error Abort Error Error Error T O T O T O T O T O Abort Aborts the communicat...

Page 76: ...ta to B by request from equipment B Request to Send Accept to Send Accept Data Data Accept Error Abort Error Abort Error Abort Error Error Error T O T O T O T O T O Abort T O Error T O Error Abort Acc...

Page 77: ...3 Open the Property in My PC Right click Computer on the start menu then select Properties 4 Click Device Manager The new dialog box is displayed The pre installed standard Windows USB driver usbser s...

Page 78: ...n Procedure Windows 7 MLM26 RTZ002F 5 Double click Unknown device The new dialog box is displayed 6 Click the Update Driver button The new dialog box is displayed 7 Click Browse my computer for driver...

Page 79: ...f file and click the OK button Normally specify the E002 folder on the distribution media The Browse for the Folder dialog box closes 10Click the Next button The Windows Security dialog box is display...

Page 80: ...Some procedures might not be displayed due to security setting 1 Connect the LM 1800P 1800PD and computer with the USB cable 2 Turn on power to the LM 1800P 1800PD 3 Point the upper right corner or lo...

Page 81: ...Installation Procedure Windows 8 MLM26 RTZ002F 5 6 Click Device Manager The new dialog box is displayed 7 Double click Unknown device The new dialog box is displayed 8 Click the Update Driver button T...

Page 82: ...wse For Folder dialog box is displayed 11Specify the location of the comclass inf file and click the OK button Normally specify the E002 folder on the distribution media The Browse For Folder dialog b...

Page 83: ...79 INSTALLATION OF USB DRIVER Installation Procedure Windows 8 MLM26 RTZ002F 5 15Click the Close button 16Turn off power to the LM 1800P 1800PD The installation of the USB driver is complete...

Page 84: ...ome procedures might not be displayed due to security setting 1 Connect the LM 1800P LM 1800PD and computer with the USB cable 2 Turn on power to the LM 1800P LM 1800PD The operating system recognizes...

Page 85: ...river software advanced The new dialog box is displayed 7 Click the Browse button The Browse for the Folder dialog box is displayed 8 Specify the location of the comclass inf file and click the OK but...

Page 86: ...ation Procedure Windows Vista MLM26 RTZ002F 10Click the Install button The Update Driver Software dialog box is displayed 11Click the Close button 12Turn off power to the LM 1800P LM 1800PD The instal...

Page 87: ...ay operating system recognizes the USB device The Welcome to the Found New Hardware Wizard dialog box is displayed 3 Select No not this time and click the Next button The new dialog box is displayed 4...

Page 88: ...dia The Browse for Folder dialog box closes 7 Click the Next button The Hardware Installation dialog box is displayed 8 Click the Continue Anyway button The Completing the Found New Hard ware Wizard d...

Page 89: ...explains the method of receiving measurement data from the LM 1800P LM 1800PD using this LAN interface Basic communication specifications 6 1 2 Cable connection Connect the LAN cable to the LAN cable...

Page 90: ...e read position using the barcode scanner magnetic card reader The ID used for display or data transmission on the device consists of up to 16 digit characters displayed on the READER screen set by th...

Page 91: ...rator of the facility Server Hub Hub Hub Hub ARK 530A NT 530 Other examination room operating room or such Connected devices Slit lamp Clinic IP ADDRESS 10 50 XXX YYY SUBNET MASK 255 255 255 0 XXX YYY...

Page 92: ...nsmitted from the LM 1800P LM 1800PD The shared folder folder path set up here is set in the parameter of the device and software such as NAVIS of the receiver computer L M X Transmission of measureme...

Page 93: ...M folder and select Share this folder under the Sharing tab The illustration to the right is an example from Windows XP 3 Click the Permissions button to display the Permis sions for LM dialog box and...

Page 94: ...irm the computer name and domain using System Properties of the computer Shared folder name see 6 3 2 Setting up the shared folder Page 88 for setting DHCP server use authorization Consult the network...

Page 95: ...rameter to Sending only 3 Set the File format parameter to NIDEK_V1 00 4 Set the DHCP parameter to Off For cases where an IP address is automatically assigned use the DHCP parameter select YES When an...

Page 96: ...d in the full view field it indicates that the space is input 5 Press the return button 6 When the confirmation screen for data saving is displayed press the SAVE button to return to the parameter scr...

Page 97: ...he shared folder LM Example 3 LM DATA A Saved in the subfolder A of the subfolder DATA of the shared folder LM 13Select the Network timeout parameter and set the timeout time after which the file save...

Page 98: ...y be incorrect Check the parameter settings and set them again correct ly Message Description DHCP error No response from the DHCP server Check the DHCP server or set the IP address subnet mask and de...

Page 99: ...arding network communication that is displayed while the device is in use Disk full No free drive space Network test completed successfully The LAN communication is confirmed being performed correctly...

Page 100: ...here is one type of output file normally produced measurement data When no style sheet file is found in the LM folder a style sheet file is also output When the style sheet file already exists a new f...

Page 101: ...nt data is no lon ger present On the receiving end after a data file is produced the existing data is read out and deleted or renamed The device indicates an error if the XML file remains present afte...

Page 102: ...st near power D 2nd near power D Prism power Prism base direction Horizontal prism power Vertical prism power UV transmittance Confidence index V1 01 only Error information V1 01 only R data Right eye...

Page 103: ...ng 64 characters LM 1800PD MachineNo Model distinction number String 64 characters ROMVersion Device ROM version for distinction String 64 characters 1 10 Version XML specification version fixed Strin...

Page 104: ...it deg Cylinder axis int 180 SE unit D Spherical equivalent value D Double 3 50 ADD unit D 1st add power D Double 1 00 ADD2 unit D 2nd add power D Double 1 50 NearSphere unit D 1st near power D Double...

Page 105: ...PrismY unit pri base up Vertical prism power base up and base down only Double 0 00 UVTransmittance unit UV transmittance int 5 PD Pupillary distance Distance unit mm Far pupillary distance Double 64...

Page 106: ...y Example Tag name Description Data type Necessity Example Opthalmology Indicates the opthalmology data Tag name Description Data type Necessity Example Common Common tag Company Company name String 6...

Page 107: ...crement Double 0 25 PrismBaseStep unit deg Prism base direction unit increment int 1 PrismMode Prism mode String pb and xy only xy AddMode Add power moder String add and nsph only add LM LM measuremen...

Page 108: ...denceIndex Confidence index String 64 characters R Right eye lens measurement value when an error occurs Error Error information String 64 characters L Left eye lens measurement value Sphere unit D Sp...

Page 109: ...ble 95 R Right eye lens measurement value Length unit mm Progressive length mm int 12 ChannelWidth unit mm Channel width mm int 6 ChannelLength unit mm Channel width position mm int 8 Index Refractive...

Page 110: ...step 0 25D Axis step 1 Cylinder mode mix Prism diopter step 0 25 Prism base step 1 Prism mode xy Add mode add Sphere Cylinder Axis SE ADD ADD2 NearSphere NearSphere2 Prism Prism base Horizontal prism...

Page 111: ...here2 Prism Prism base Horizontal prism Vertical prism UV Transmittance ConfidenceIndex 1 00D 0 75D 84 2 75D 1 75D 1 75 273 out0 00 down1 75 Distance Distance right Distance left 62 0mm 32 0mm 30 0mm...

Page 112: ...No ID pati ID FirstName FirstName MiddleName MiddleName LastName LastName Sex Sex Age Age DOB DOB NameJ1 NameJ1 NameJ2 NameJ2 Patient Patient tag RT ID or print No Patient ID Patient first name Patien...

Page 113: ...arSphere2 Prism unit pri 2 25 Prism PrismBase unit deg 77 PrismBase PrismX unit pri base in 0 50 PrismX PrismY unit pri base up 2 00 PrismY UVTransmittance unit 5 UVTransmittance R Right eye lens meas...

Page 114: ...ChannelWidth unit mm 18 ChannelWidth ChannelLength unit mm 23 ChannelLength Index Index R Right eye lens measurement value Progressive length mm Channel width mm Channel width position mm Refractive...

Page 115: ...stName LastName Sex Sex Age Age DOB DOB NameJ1 NameJ1 NameJ2 NameJ2 Patient Patient tag RT ID or print No Patient ID Patient first name Patient middle name Patient last name Patient sex Patient age Pa...

Page 116: ...base direction Horizontal prism power Vertical prism power UV transmittance Confidence index S Error MEASUREMENT ERROR Error S Single state measurement value when an error occurs Error information R S...

Page 117: ...eye far pupillary distance NIDEK LM measurement data NIDEK inherent optical tag S Length unit mm 23 Length ChannelWidth unit mm 18 ChannelWidth ChannelLength unit mm 23 ChannelLength Index Index Gree...

Page 118: ...Data Format MLM26 RTZ002F Inside InsideR unit mm 2 0 InsideR InsideL unit mm 1 0 InsideL Inside Inside amount Right eye lens near portion inside amount Left eye lens near portion inside amount NIDEK...