7. Maintenance
NIC-TD-0000025-00
- 52
–
7.2.1.
Exchange Lamp
Displays when the lamp ON time exceeds 2,000 hours.
Switch off the equipment and replace the lamp as soon as possible.
Method of handling the warning
After pressing the "OK" button on the Warning screen return to the READY screen before
turning off the power switch.
*
If a measurement is in progress terminate it before switching off the equipment. Failure
to do so could result in the data not being recorded.
<Warning screen>
<READY screen>
Use of a deteriorated lamp can result in increasing the margin of error of the measured
value.
Contact one of our service representatives as soon as possible. (Refer to 10 "Contact".)
7.2.2.
Exchange Pump
Displays when the pump ON time exceeds 2,000 hours.
Switch off the equipment and replace the pump as soon as possible.
Method of handling the warning
After pressing the "OK" button on the Warning screen return to the READY screen before
turning off the power switch.
*
If a measurement is in progress terminate it before switching off the equipment. Failure
to do so could result in the data not being recorded.
<Warning screen>
<READY screen>
Use of a badly deteriorated pump can result in increasing the margin of error of the
measured value.
Contact one of our service representatives as soon as possible. (Refer to 10 "Contact".)