7. Maintenance
NIC-TD-0000025-00
- 54
–
7.2.4.
Battery Empty
Displays when the remaining battery level is too low.
If the equipment continues to be used with a low battery, its power will turn off within 10
minutes.
Method of handling the warning
After pressing the "OK" button on the display screen, return to the READY screen.
If a measurement is in progress, terminate immediately it to record the data.
<Warning screen>
<READY screen>
After returning to the READY screen, turn off the power to the equipment, and then remove
the cover at the bottom of the equipment. Replace the battery with a fully charged one and
power on the equipment again to enable measurements to be made.