Chapter 8. Counting and Analysis Workflow
8001009 Cellometer® Auto T4 User Manual Rev C
38
CALCULATE SAMPLE ADJUSTMENT
The
Sample Adjustment
tool assists you in calculating the adjustment necessary in order to achieve a specified
target concentration or target number of cell
s
in a tube. Before calculating this adjustment, you must first perform
a count on either a live image or saved image to display the current cell count and concentration (or live/dead
count, total cell count, viability %, live cell concentration and total cell concentration if
Test Viability
was selected).
To calculate sample adjustment:
1.
Click on the
Sample Adjustment
icon to launch the tool.
2.
Choose the concentration source to be used for calculating the
adjustment (live, dead or total concentration) and click
Continue
.
3.
Enter the
Original Sample Volume
. The
Total Cell Number in
Sample
value will be updated according to sample volume.
4.
Choose whether you want a
Target Concentration (cells/mL)
or a
Target Number of Cells
and enter the target concentration
or cell count accordingly.
5.
Click
Apply Change
. In the
Sample Adjustment
field, a statement will appear indicating how to adjust the
sample to reach the targeted concentration or cell count.
6.
Click the
Print with Report
check box to include adjustment results in the data report.
Alternatively, this tool can be activated by selecting the
Sample Adjustment
option from the Analysis Menu.
Summary of Contents for Cellometer Auto T4
Page 1: ......
Page 6: ...Table of Contents 8001009 Cellometer Auto T4 User Manual Rev C vi ...
Page 12: ...Chapter 1 Introduction 8001009 Cellometer Auto T4 User Manual Rev C 6 ...
Page 16: ...Chapter 2 Equipment Safety 8001009 Cellometer Auto T4 User Manual Rev C 10 ...
Page 26: ...Chapter 5 Operation 8001009 Cellometer Auto T4 User Manual Rev C 20 ...
Page 36: ...Chapter 6 Software GUI Elements 8001009 Cellometer Auto T4 User Manual Rev C 30 ...
Page 46: ...Chapter 8 Counting and Analysis Workflow 8001009 Cellometer Auto T4 User Manual Rev C 40 ...
Page 56: ...Chapter 9 Cell Types 8001009 Cellometer Auto T4 User Manual Rev C 50 ...
Page 66: ...Chapter 13 Nexcelom Support 8001009 Cellometer Auto T4 User Manual Rev C 60 ...
Page 70: ...Appendix A Consumables 8001009 Cellometer Auto T4 User Manual Rev C 64 ...
Page 72: ...Appendix B Declaration of Conformity 8001009 Cellometer Auto T4 User Manual Rev C 66 ...