7
CAUTION
When lowering the table on to the supports, do not allow the
table to shift sideways. Doing so could damage the isolators
and cause degradation in performance.
2.2
Leveling the Table
Use the following procedure to level the table. The special wrench (Figure 3)
is for all adjustments.
Figure 3
— Isolator Leveling Wrench
1.
Refer to Figure 4. As shown, use one of the special wrenches to hold
the hex nut stationary. Insert the prongs of the second wrench into the
holes on the bottom of the support plate. Rotate the plate using the top
wrench to adjust the height.
2.
Place a level on the table top, approximately in the center.
3.
For each isolator, adjust the support plates as required to bring the table
to a level condition. Be sure to check the level using both the length and
width axes of the table. Verify that the centering disks are centered in
and resting on their respective guides (Figure 4).
Figure 4
— Level Adjustment of Isolators