Logging Data to a File for Upload to PC
1919-R User Manual
83
Store By:
Determines the quality of the data that is stored. Set logging to store the
average of several readings or only one of the group. (If
Sample Rate
is set to
Every
Reading
, then this parameter is disabled and grayed out.) Options include:
Sample:
Logs only the last reading that was measured. (Reduces log file size by storing
only one of several measurements.)
Average:
Stores the average of all of the readings that have been measured since the
previous data has been added to the log.
10.3
Start Logging
You activate logging by pressing the
Start
key in the
Log Setup
screen.
To start logging:
1.
In the
Log Setup
screen, press the
Start
key. The main measurement screen displays the graph
type that was previously selected in the Display menu, and the following additional
parameters at the top of the screen:
Elapsed
: Time that has elapsed since starting the present log session (in the example
below, 3 minutes).
Readings
: Number of measurements logged to the file since the start of the present log
session (in the example below, 906).
Stop
: Condition for ending the present log session (in the example below, after storing
20000 readings).
Figure
10-11 Started Logging
2.
While logging is active, to stop the log manually, press
Stop
. This stops the logging
immediately and saves the date. (Press
Abort
to abort the log, exit the logging screen, and
revert back to the
Log Setup
Screen.)
3.
When the log is completed, the display will stop updating and a Log Summary appears. The
following will be displayed in a message box in the display area:
Log file name
Total readings
Number of overrange
Maximum
Minimum
Average