Calibration, Maintenance, and Troubleshooting
1919-R User Manual
103
factor for each pulse width option available in the sensor. For sensors with a removable
diffuser, there are two sets of factors; one set for each diffuser setting.
Figure
12-26 Calibration Activated on Main Measurement Screen
4.
Press
Modify
. The screen displays the
Original
(current) reading and factor (on the left side of
the screen), and
Updated
factor and its effect on the reading (on the right). The
Updated
box
is highlighted, indicating it is modifiable.
Figure
12-27 Modifying Calibration for Pyroelectric Sensors
5.
In the
Updated
area, use the Right/Left arrows to select the digit to update. Use the Up/Down
arrows to change the value of the selected digit.
6.
Press
Save
to save the settings and return to the main Measurement screen, which now
displays the updated Overall Factor.
Figure
12-28 Updated Overall Factor for Pyroelectric Sensors
7.
Press
Exit
to return to the Menu screen.