Calibration, Maintenance, and Troubleshooting
1919-R User Manual
97
5.
Press
Next
and then unblock the laser.
Figure
12-15 Updated Response Calibration Factor in Main Screen
6.
Press
Next
. The calibration is updated in the main Measurement screen.
Initially, the screen displays the updated
Scaled
,
Present
, and
Response Factor
values.
Figure
12-16 Scaled, Present, and Response Factor Values Updated
Then, the screen displays the final
Response Factor
value.
Figure
12-17 Final Response Calibration Factor Value
The graph (as shown above) is drawn with two dotted lines; one at 100% of the expected
measurement at one at 90%. 100% is drawn at about 75% of the full height of the graph to
allow the user to see the overshoot, if there is any.
7.
Press
Finish
to save the Response factor.
8.
Press
Back
to change and test again. (Repeat steps 5-7 above.)