1919-R User Manual
77
10
Logging Data to a File for Upload to PC
In addition to displaying the results visually on the screen, you may need to gather a larger amount
of data and analyze it in a more thorough fashion. 1919-R provides the ability to store the data and
review it with Newport’s PMManager application, Excel, or another application that better fits
your needs.
This section describes how to log laser measurement data to a file stored on an external storage
device for later upload to a PC.
Topics include:
Setting Logging Parameters for the Various Stop Modes
Viewing the Log File on Your Computer
10.1
Launching the Log Setup
To activate the Log function, you need to first connect an external storage device to the 1919-R
meter. Once connected, the
Log Setup
screen is displayed, where you can configure the logging
parameters and start the logging process.
To launch the Log Setup:
1.
Insert an external storage device Disk on Key into the USB port located on the left side of the
1919-R meter.
2.
Press
Menu
and select
Log
.
Upon pressing
Log
in the
Menu
screen, 1919-R will search for a Disk on Key. If none is
attached, it will display an error message, “External storage device not available.”
If a Disk on Key is found, 1919-R will look for a directory called “Log Files” in the root directory
of the Disk on Key, and if not found, 1919-R will create it. 1919-R will then set “Log Files” as
the active directory. 1919-R will not allow browsing of any other directories on the Disk on
Key.
Figure
10-1 Log Files Directory on Disk On Key