51
Specifications
Table 9-1 (continued)
Model
XP-3360II
XP-3368II
Target gas
Combustible gas/
solvent gas
Combustible gas/
solvent gas
Oxygen
Detection principle
Catalytic
Catalytic
Galvanic cell
Measuring range
(Service range)
0–5000 ppm
(5001–5500 ppm)
0–5000 ppm
(5001–5500 ppm)
0–25vol%
(25.1–50.0vol%)
0–10000 ppm
(10001–11000 ppm)
0–10000 ppm
(10001–11000 ppm)
Accuracy
10%F.S.: ±1%F.S.
10%F.S.: ±5%F.S.
10%F.S.: ±1%F.S.
10%F.S.: ±5%F.S.
±0.3vol%
Resolution
1 ppm
1 ppm
0.1vol%
Gas alarm set value
250 ppm or 500 ppm
250 ppm
or
500
ppm
18vol%
Response time
T90: 30 seconds
T90: 30 seconds
–
Continuous operation time
Approx. 15 hours
Approx. 15 hours
F.S.: Full scale
Model
XP-3340II
XP-3380II
XP-3380II-E
Target gas
Combustible gas/
solvent gas
Oxygen
Oxygen in
Detection principle
Thermal
conductivity
Galvanic cell
Measuring range
(Service range)
0–100vol%
(100.1–110.0vol%)
0–25vol%
(25.1–50.0vol%)
Accuracy
±5%F.S.
±0.3vol%
Resolution
0.1vol%
0.1vol%
Gas alarm set value
50vol%
18vol%
Response time
T90: 60 seconds
-
Continuous operation time
Approx.100 hours
Approx.100 hours