15
Operation
Press and hold the power button until the power-on screen (company logo ) is
displayed on the LCD.
--> After the unit beeps once, the “model, serial No. and battery type”, “target gases”,
“full scale values”, “alarm set values”, “previous calibration date” and “log memory
usage”, and
“
Bluetooth not available
”
screens will be displayed in sequence. When
the unit gives off a beep, a warm-up cycle will start. The warm-up progress bar will be
displayed. When the warm-up cycle is completed, “Warm-up completed 100%” will be
displayed and a zero adjustment will be automatically started. When the zero
adjustment is completed, the unit will give a long beep and the gas concentrations
screen (HOME) will be displayed.
Power on Model, serial No., battery type
Target gases Full scale values
Alarm set values Previous calibration date Log memory usage Bluetooth not available
Other than XP-3340II
XP-3340II
Zero adj. completed
HOME
Warm-up cycle
Warm-up cycle takes a maximum of 5 minutes.
Only Gas1 and its information is displayed for a single-gas
detector.
Only the power off button is enabled during warm-up cycle.
Operation of other buttons is disabled.
If an error message appears, see “Error Messages” on pages 36
and 37.
The previous calibration date screen is skipped when shipped,
because there is no record.
The log memory usage screen is skipped for the unit with no
Bluetooth function installed.
The Bluetooth not available screen is skipped for the unit with
Bluetooth function installed.