30
Settings
The factory default of the logging interval rate is one second.
The minimum settable interval rate is 200 ms.
The logging period is approx. 40 hours when the logging interval
rate is set to 10 seconds for 2-gas models. It is approx. 70 hours
for 1-gas models.
Up to 500 log files can be saved; more than 500 files cannot be
saved even if free space is available.
If the device clock is inaccurate, accurate logging is not possible.
Adjust the clock. (Clock adjustment on page 33)
Logging will automatically stop if:
-
the detector powers off (logged data will be automatically
saved),
-
log memory usage (used space) reaches 100%, or
-
an error occurs, except for a pump error (logging will continue
even when a pump error occurs).
Log data readout requires a PC + a data logger software (XP-
3000IIL, separately sold) or an Android (ver. 7.0 or higher) or iOS
(ver. 10.0 or higher) device with built-in Bluetooth. For the
readout procedure, see the data logger software’s instruction
manual or follow the instructions in the free app “Utility for gas
detector”.
If the power is interrupted for a few seconds (e.g., by the detector
being bumped or dropped) while logging, the logs collected one
minute before the interruption will not be saved.
The logged gas concentrations within the full-scale range (not
service range) can be viewed on a cell phone or PC.
NOTE