17
9.0
Program Information
Program
No.
1
2
3
4
5
Frequency
(Hz)
4
35
99
Channel 1: 4
Channel 2: 99
4 - 99
Treatment
Time (min)
open
open
open
open
open
Pulse duration
(
µ
sec)
150
150
150
150
150
Burst or Trigger
High Frequency Trigger
Burst - Frequency of 99Hz, pulse width 120
µ
sec. for
3 sec.; Will return to program settings for 1.5 sec.
Burst - Frequency of 4Hz, pulse width 120
µ
sec. for 3
sec.; Will return to program settings for 3 sec.
Trigger
Trigger
Indications
Chronic Pain
Joint Pain
Labour Pain (Trigger)
Acute Pain
Acute Pain
Pain Block
Chronic & Acute Pain
Arthritic Pain
Severe Acute Pain
Arthritic Pain
Period Pain
PROGR
AM
INF
ORMA
TION
Program 1 (High Frequency Trigger):
Program 1 delivers a stimulus of 4Hz. When the Trigger button is pressed a high frequency 99Hz stimulus is delivered. To disable High
Frequency Trigger mode press the Trigger button a second time.
Program 2, 3 (Burst Mode):
Burst mode is available on Programs 2 and 3. When the Burst button is pressed Burst mode is enabled. To disable Burst mode press
the button a second time. If Burst mode is selected during Program 2, it results in a frequency of 99Hz at a pulse width of 120
µ
s for 3
seconds. This then returns to the normal program frequency and pulse width for 1.5 seconds. If Burst mode is selected during Program
3, it results in a frequency of 4 Hz at a pulse width of 120
µ
s for 3 seconds, before returning to the normal program frequency and
pulse width for 3 seconds.
Program 4 (Trigger) Dual Frequency:
When program 4 is selected channel 1 and channel 2 deliver stimuli at different frequencies. Channel 1 delivers a stimulus of 4 Hz while
channel 2 delivers a stimulus of 99Hz.
Program 5 (Trigger) Massage:
Program 5 is a variable program. This delivers a stimulus of 4 to 99Hz.
Trigger mode is available on Program 4 and 5. When the Trigger button is pressed Trigger mode is enabled. When the button is pressed
a second time the unit enters a contraction cycle for as long as the button is pressed. When the button is released the unit enters the
relaxation cycle. When Trigger mode is disabled, by pressing the intensity button, the stimulation builds over a 2-second period to the
previously set amplitude level.
English