14. If you want to interrupt the treatment session (e.g. to reposition the electrodes), briefly press the on/off
button. The unit issues a beep and the pause icon appears on the screen (Display 7). To deactivate the pause
function, press the on/off button again. Then the treatment session is restarted from where it was paused
and the pause icon disappears from the display.
15. The
Burst mode
( ) can be used during Programs 2 and 3. Pushing the button enables Burst mode.
To deactivate the function, press the button again (Display 8).
The
Trigger mode
( ) can be used during Programs 1, 4, and 5. When the Trigger button is pressed
while in program 1, the signal frequency increases from 4Hz to 99Hz. Press the Trigger button a second
time to disable Trigger mode. When the Trigger button is pressed once during program 4 or 5, Trigger
mode is enabled. When the button is pressed a second time the unit enters a contraction cycle which lasts
as long as the button remains pressed. When the button is released the unit enters the relaxation cycle.
Trigger mode is disabled by pressing the intensity button and the stimulation builds over a 2-second period to
the previously set amplitude level
16. MediTens XP has a load sense function that monitors the connection between the lead and its electrode
and between the electrode and the skin. When poor skin contact is detected (Display 9):
•
The intensity bars of the channel affected will flash.
•
The warning icon ( ) flashes on the display.
•
An audible beep will emit from the unit.
•
The treatment session timer pauses.
•
The intensity value falls to zero and the intensity increase button is deactivated.
When proper contact is restored, stimulation builds over a 2-second period to the previously set intensity level.