NED
UME-0034-03
XCM8060SA_8040SA_6040SA
29
4.2 Details on Commands
4.2.1 Setting Analog Gain
Sets analog gain in 21 steps between x1 and x11.2. (See Table 4-9)
Format 2
CMD
□
VAL1 CR
CMD
gax
VAL
0 (x1)
–20 (x11.2)
<Example>
gax
□
5 CR (Setting analog gain 5(x1.84))
>OK
>gax 5
4.2.2 Setting Digital Gain
Sets digital gain in 512 steps between x1 and x2.
Format 2
CMD
□
VAL1 CR
CMD
gdx
VAL
0(x1) - 511(x2)
<Example>
gdx
□
255 CR (Setting digital gain 255(1023/(1023-255)=x1.33))
>OK
>gdx 255
4.2.3 Setting Digital Offset
Sets digital offset -15 to +15(8 bit:1DN/Step), -60 to +60(10 bit:4DN/step)
Format 2
CMD
□
VAL1 CR
CMD
odx
VAL
-15 to +15
<Example>
odx
□
5 CR (Setting digital offset 5(8-bit) or 20(10-bit))
>OK
>odx 5