Trex™ HD Monitoring Amplifier
User & Service Manual
10
2.1.2. Declaration of Compliance for IEC 60601-1-2
Table 1 - Electromagnetic Emissions
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The
Trex HDTrex HD Monitoring
is intended for use in the electromagnetic environment specified
below. The customer or the
Trex HD Monitoring
should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment – Guidance
RF emissions
CISPR 11
Group 1
The Trex HD Monitoring uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Trex HD Monitoring is suitable for use in all
establishments, including domestic establishments and
those directly connected to public low voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
1
The conducted, harmonic, and flicker emissions testing was only performed on the mains-powered
camcorder and mains-powered PC when providing power to the Trex HD Monitoring amplifier via USB
connection. This testing was not performed on the Trex HD Monitoring amplifier in isolation because it is
a battery-powered device.