background image

Overview

This manual provides the necessary information to maintain and service the
Olympic Bili-Meter

TM

. The operating instructions in this manual are intended for 

use under the direct supervision of a licensed medical practitioner. The service
instructions in this manual are intended for use by qualified technicians.

Conventions

The following conventions are used in this manual.

Table 1

Conventions

Symbols

The following symbols are located on the Olympic Bili-Meter and its packaging.

Table 2

Symbols

OLYMPIC BILI-METER

Instruction

Manual

1

Symbol

Definition

Shipping

Storage

Temperature

Manufacturer

Symbol

Definition

Atmospheric pressure

Battery

Caution

Humidity,

condensing

Convention

Description

NOTE

Notes provide additional information to clarify a point in the

text.

CAUTION

Cautions indicate situations that, if not avoided, could result in 

minor

to moderate injury to the patient or operator, or damage to the

equipment.

WARNING

Warnings indicate situations that, if not avoided, could result in 
serious injury or death to the patient or opera

tor.

BUTTON

This character style represents buttons and controls that the user 
can touch or

press.

Summary of Contents for OLYMPIC BILI-METER 22

Page 1: ...nnecting the Sensor 4 Taking Measurements 5 Cleaning 7 Maintenance and Service 7 Replacing the Battery 7 Calibrating the Bili Meter 8 Service and Repair 8 Returning for Calibration or Service 8 Contac...

Page 2: ...The information in this manual is subject to change without notice No part of this manual may be photocopied reproduced translated or reduced to any electronic medium without the express written perm...

Page 3: ...its packaging Table 2 Symbols OLYMPIC BILI METER Instruction Manual 1 Symbol Definition Shipping Storage Temperature Manufacturer Symbol Definition Atmospheric pressure Battery Caution Humidity conde...

Page 4: ...22 readout Sensor cable connection Display ON READ button Shown with Type B 22 sensor Type B 22 Sensor This sensor is used to measure the irradiance of phototherapy lights placed over the patient e g...

Page 5: ...are centimeter per nanometer W cm2 nm A nanometer is a measure of wavelength equal to one billionth of a centimeter The term per nanometer indicates the average irradiance per nanometer across the spe...

Page 6: ...ert the connector into the socket Figure 5 Connecting the sensor to the readout Socket Readout Connector Figure 6 Disconnecting the sensor Pull Readout Connector sleeve 4 OLYMPIC BILI METER Instructio...

Page 7: ...taken Figure 7 Measuring with the Type B 22 sensor Phototherapy light Sensor aimed for maximum reading 3 Press and hold READ On briefly displays then the irradiance measurement displays While pressing...

Page 8: ...Messages The following messages may show on the display of the readout Table 8 1 Messages 6 OLYMPIC BILI METER Instruction Manual Display Message Description Low Bat Battery power is low replace the...

Page 9: ...he back cover of the readout see Figure 9 An instruction label on the inside of the back cover illustrates how to remove and insert the battery NOTE Figure 9 Replacing the battery Back cover Screws x4...

Page 10: ...ts and must be returned to Natus Medical for all repairs and parts replacement After any service the Bili Meter must be recalibrated radiometrically to assure accurate measurement of irradiance Return...

Page 11: ...sor 53822 Bili Meter Model 22 with Type F 22 sensor 53825 Type B 22 sensor only 53860 Type F 22 sensor only 53865 Note The Type B 22 and Type F 22 sensors can be used interchangeably with the Bili Met...

Page 12: ...ields greater than or equal to 1 V m or contact electrostatic discharge abnormal operation eg display fluctuation or lock up may occur If so re orient the meter and or allow it to auto reset and retak...