background image

Taking

Measurements

To use the Type B-22 sensor:

1

Connect the sensor to the readout.

2

Hold the sensor against the infant’s body as near to the umbilical as possible 

and aim the sensor at the center of the phototherapy light (see Figure 7).

For reproducible measurements, always hold the sensor on the same place on 
the infant’s body.

Changes in the distance or angle of the light to the patient will change the 
irradiance the patient receives, requiring new measurements to be taken.

Figure 7

Measuring with the Type B-22 sensor

Phototherapy

light

Sensor aimed for maximum reading

3

Press and hold

READ

. “On” briefly displays, then the irradiance measurement 

displays. While pressing

READ

, adjust the aim of the sensor to obtain the 

maximum reading.

Release

READ

. The irradiance measurement locks on the display for 30 seconds 

then automatically shuts off to preserve battery life.

4

When charting irradiance, note the type of radiometer used (e.g., Olympic
Bili-Meter) and sensor (e.g., Type B-22).

If more than one phototherapy light is being used on the infant, take separate 
measurements for each light and chart each reading.

OLYMPIC BILI-METER

Instruction

Manual

5

Summary of Contents for OLYMPIC BILI-METER 22

Page 1: ...nnecting the Sensor 4 Taking Measurements 5 Cleaning 7 Maintenance and Service 7 Replacing the Battery 7 Calibrating the Bili Meter 8 Service and Repair 8 Returning for Calibration or Service 8 Contac...

Page 2: ...The information in this manual is subject to change without notice No part of this manual may be photocopied reproduced translated or reduced to any electronic medium without the express written perm...

Page 3: ...its packaging Table 2 Symbols OLYMPIC BILI METER Instruction Manual 1 Symbol Definition Shipping Storage Temperature Manufacturer Symbol Definition Atmospheric pressure Battery Caution Humidity conde...

Page 4: ...22 readout Sensor cable connection Display ON READ button Shown with Type B 22 sensor Type B 22 Sensor This sensor is used to measure the irradiance of phototherapy lights placed over the patient e g...

Page 5: ...are centimeter per nanometer W cm2 nm A nanometer is a measure of wavelength equal to one billionth of a centimeter The term per nanometer indicates the average irradiance per nanometer across the spe...

Page 6: ...ert the connector into the socket Figure 5 Connecting the sensor to the readout Socket Readout Connector Figure 6 Disconnecting the sensor Pull Readout Connector sleeve 4 OLYMPIC BILI METER Instructio...

Page 7: ...taken Figure 7 Measuring with the Type B 22 sensor Phototherapy light Sensor aimed for maximum reading 3 Press and hold READ On briefly displays then the irradiance measurement displays While pressing...

Page 8: ...Messages The following messages may show on the display of the readout Table 8 1 Messages 6 OLYMPIC BILI METER Instruction Manual Display Message Description Low Bat Battery power is low replace the...

Page 9: ...he back cover of the readout see Figure 9 An instruction label on the inside of the back cover illustrates how to remove and insert the battery NOTE Figure 9 Replacing the battery Back cover Screws x4...

Page 10: ...ts and must be returned to Natus Medical for all repairs and parts replacement After any service the Bili Meter must be recalibrated radiometrically to assure accurate measurement of irradiance Return...

Page 11: ...sor 53822 Bili Meter Model 22 with Type F 22 sensor 53825 Type B 22 sensor only 53860 Type F 22 sensor only 53865 Note The Type B 22 and Type F 22 sensors can be used interchangeably with the Bili Met...

Page 12: ...ields greater than or equal to 1 V m or contact electrostatic discharge abnormal operation eg display fluctuation or lock up may occur If so re orient the meter and or allow it to auto reset and retak...

Reviews: