Cleaning
Required items:
Soft cloth
Mild cleaning detergent (e.g., Virex
TM
Tb, Virustat
®
, Coverage
®
) or
mild soap-and-water solution
To clean the
Bili-Meter:
1
Confirm that the Bili-Meter is off. If it is on, press
POWER
to turn it off.
2
Dampen the cloth with either the mild soap-and-water solution or detergent,
then wipe down all exterior surfaces of the Bili-Meter.
Maintenance and Service
CAUTION
this
device.
Replacing the Battery
Required
items:
5/64-in. FLS Allen wrench
9-volt alkaline, non-rechargeable battery
To replace the battery:
1
Remove the back cover: Loosen and remove the four screws that secure the
back cover of the readout (see Figure 9).
An instruction label on the inside of the back cover illustrates how to remove and insert
the battery.
NOTE
Figure 9
Replacing the
battery
Back
cover
Screws
(x4)
Battery
Polarity
icons
Battery
holder
2
Insert the new 9-volt battery into the battery holder (see Figure 9).
OLYMPIC BILI-METER
Instruction
Manual
7
Only qualified technicians should maintain or service this
device.
Read and be familiar with this instruction manual before maintaining or servicin
g
The printed circuit boards
(PCBs) contain static sensitive parts. Always use appropriate
electrostatic discharge protection, such as an electrical-grounding wrist strap, when
working with internal
components.
To assure accuracy, factory-calibrate the Bili-Meter to a radiometric standard annually
(see page 8).