To use the Type F-22 sensor:

1

Connect the sensor to the readout.

2

Place the sensor, with white diffuser down, in the middle of the lighted side of 

the pad or mattress (see Figure 8). Any disposable covers should be on the pad 
or mattress.

Figure 8

Measuring with the Type F-22 sensor

Type F-22 sensor

Phototherapy pad, lighted side 

up 

with pad cover in

place

3

Press and hold

READ

. “On” briefly displays, then the irradiance measurement 

displays.

Release

READ

. The irradiance measurement locks on the display for 30 seconds 

then automatically shuts off to preserve battery life.

Take two additional readings at different places on the pad, then average all 
three readings. If a pad is being used at the same time as an overhead 
phototherapy light, measure and chart the irradiance of the pad and the light 
separately with the appropriate type of sensor.

4

5

Messages

The following messages may show on the display of the readout.

Table 8.1

Messages

6

OLYMPIC BILI-METER

Instruction

Manual

Display

Message

Description

Low

Bat

Battery power is low; replace the battery

(see page 7).

Over Range

If

150 

flashes on the display when

READ

is pressed,

the 

irradiance is greater than 150 µW/cm²/nm and is out of range

of 

the

Bili-Meter.

Error

If dashes flash on the display when

READ

is pressed, an error 

condition is present:

The sensor may not be connected. Check the sensor 
connection and verify the sensor version.
The wrong version of sensor may be connected to the

digital 

readout unit. The color of the sensor and readout

calibration 

labels must

match.

The sensor may have failed. Retry the read. If the

problem 

continues, contact Natus Medical

(see page 8).

The digital readout may have failed. Retry the read. If

the 

problem continues, contact Natus Medical

(see page 8).

Always make sure that the entire white diffuser of the sensor is over the lighted area of

the

CAUTION

pad, otherwise inaccurate measurements may r

esult.

Summary of Contents for OLYMPIC BILI-METER 22

Page 1: ...nnecting the Sensor 4 Taking Measurements 5 Cleaning 7 Maintenance and Service 7 Replacing the Battery 7 Calibrating the Bili Meter 8 Service and Repair 8 Returning for Calibration or Service 8 Contac...

Page 2: ...The information in this manual is subject to change without notice No part of this manual may be photocopied reproduced translated or reduced to any electronic medium without the express written perm...

Page 3: ...its packaging Table 2 Symbols OLYMPIC BILI METER Instruction Manual 1 Symbol Definition Shipping Storage Temperature Manufacturer Symbol Definition Atmospheric pressure Battery Caution Humidity conde...

Page 4: ...22 readout Sensor cable connection Display ON READ button Shown with Type B 22 sensor Type B 22 Sensor This sensor is used to measure the irradiance of phototherapy lights placed over the patient e g...

Page 5: ...are centimeter per nanometer W cm2 nm A nanometer is a measure of wavelength equal to one billionth of a centimeter The term per nanometer indicates the average irradiance per nanometer across the spe...

Page 6: ...ert the connector into the socket Figure 5 Connecting the sensor to the readout Socket Readout Connector Figure 6 Disconnecting the sensor Pull Readout Connector sleeve 4 OLYMPIC BILI METER Instructio...

Page 7: ...taken Figure 7 Measuring with the Type B 22 sensor Phototherapy light Sensor aimed for maximum reading 3 Press and hold READ On briefly displays then the irradiance measurement displays While pressing...

Page 8: ...Messages The following messages may show on the display of the readout Table 8 1 Messages 6 OLYMPIC BILI METER Instruction Manual Display Message Description Low Bat Battery power is low replace the...

Page 9: ...he back cover of the readout see Figure 9 An instruction label on the inside of the back cover illustrates how to remove and insert the battery NOTE Figure 9 Replacing the battery Back cover Screws x4...

Page 10: ...ts and must be returned to Natus Medical for all repairs and parts replacement After any service the Bili Meter must be recalibrated radiometrically to assure accurate measurement of irradiance Return...

Page 11: ...sor 53822 Bili Meter Model 22 with Type F 22 sensor 53825 Type B 22 sensor only 53860 Type F 22 sensor only 53865 Note The Type B 22 and Type F 22 sensors can be used interchangeably with the Bili Met...

Page 12: ...ields greater than or equal to 1 V m or contact electrostatic discharge abnormal operation eg display fluctuation or lock up may occur If so re orient the meter and or allow it to auto reset and retak...

Reviews: