Revised 8/27/13
4-5
Electromagnetic Compatibility (EMC)
Nicolet Cortical Stimulator
Table 3 – Guidance and manufacturer’s declaration – electromagnetic immunity – for the Nicolet Cortical
Stimulator
.
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Nicolet Cortical Stimulator is intended for use in the electromagnetic environment specified below. The customer or the user of the
Nicolet Cortical Stimulator should assure that it is used in such an environment.
Immunity test
IEC 60601test level
Compliance level
Electromagnetic environment –
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply lines
±1 kV for signal lines
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5%
U
T
(>95% dip in
U
T
) for
0.5 cycle
40%
U
T
(60% dip in
U
T
) for 5
cycles
70%
U
T
(30% dip in
U
T
) for 25
cycles
<5%
U
T
(>95% dip in
U
T
) for 5
sec
<5%
U
T
(>95% dip in
U
T
) for
0.5 cycle
40%
U
T
(60% dip in
U
T
) for 5
cycles
70%
U
T
(30% dip in
U
T
) for 25
cycles
<5%
U
T
(>95% dip in
U
T
) for 5
sec
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the Nicolet
Cortical Stimulator requires continued
operation during power mains
interruptions, it is recommended that
the Nicolet Cortical Stimulator be
powered from an uninterruptable
power supply or a battery.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
Summary of Contents for Nicolet SCU
Page 2: ......
Page 9: ...Revised 8 27 13 1 1 1 General Information...
Page 10: ...Nicolet Cortical Stimulator 1 2 Revised 8 27 13 Blank Page...
Page 21: ...Revised 8 27 13 2 1 2 Nicolet Cortical Stimulator Overview...
Page 22: ...Nicolet Cortical Stimulator 2 2 Revised 8 27 13 Blank Page...
Page 37: ...Revised 8 27 13 3 1 3 Performing a Cortical Stimulation Study...
Page 38: ...Nicolet Cortical Stimulator 3 2 Revised 8 27 13 Blank Page...
Page 55: ...Revised 8 27 13 4 1 4 Electromagnetic Compatibility EMC...
Page 56: ...Nicolet Cortical Stimulator 4 2 Revised 8 27 13 Blank Page...
Page 62: ...Nicolet Cortical Stimulator 4 8 Revised 8 27 13 Blank page...
Page 63: ...Revised 8 27 13 5 1 5 Error Codes...
Page 64: ...Nicolet Cortical Stimulator 5 2 Revised 8 27 13 Blank Page...
Page 66: ...Nicolet Cortical Stimulator 5 4 Revised 8 27 13 Blank page...