Revised 8/27/13
3-13
Performing a Cortical Stimulation Study
5.
Train Duration
- Press the
Train Duration
button and then select the
desired train duration: Units of
seconds
(5 seconds, typical) or
Continuous
if using a stimulator probe. The last selected rate setting becomes the startup
default setting.
6.
Pulse Frequency
- Press the
Pulse Frequency
button and then select the
desired pulse frequency rate (50 Hz, typical). The last selected rate setting
becomes the startup default setting.
7.
Pulse Duration
- Press the
Pulse Duration
button and then select the
desired pulse duration:
100 - 1000 uSec
. The last selected Duration setting
becomes the startup default setting.
8.
Positive Electrode
s - Press the ‘
Ele
’ (Anode) button and turn the
Selector
knob
until the desired electrode number or label appears and then
press the
Selector
knob
.
9.
Negative Electrode
s - Press the ‘
Electrode -
’ (Cathode) button and turn the
Selector
knob
until the desired electrode number or label appears and then
press the
Selector
knob
.
The applicable stimulation probe is the Natus Neurology Incorporated
Bipolar Stimulation Probe with 2.2mm diameter tips. The applicable
stimulation electrodes are any commercially available 2.3mm diameter
Strip and Grid Electrodes. In the United States only, the stimulation
electrodes and stimulation probes must have been cleared to market by
the Food and Drug Administration (FDA).
Summary of Contents for Nicolet SCU
Page 2: ......
Page 9: ...Revised 8 27 13 1 1 1 General Information...
Page 10: ...Nicolet Cortical Stimulator 1 2 Revised 8 27 13 Blank Page...
Page 21: ...Revised 8 27 13 2 1 2 Nicolet Cortical Stimulator Overview...
Page 22: ...Nicolet Cortical Stimulator 2 2 Revised 8 27 13 Blank Page...
Page 37: ...Revised 8 27 13 3 1 3 Performing a Cortical Stimulation Study...
Page 38: ...Nicolet Cortical Stimulator 3 2 Revised 8 27 13 Blank Page...
Page 55: ...Revised 8 27 13 4 1 4 Electromagnetic Compatibility EMC...
Page 56: ...Nicolet Cortical Stimulator 4 2 Revised 8 27 13 Blank Page...
Page 62: ...Nicolet Cortical Stimulator 4 8 Revised 8 27 13 Blank page...
Page 63: ...Revised 8 27 13 5 1 5 Error Codes...
Page 64: ...Nicolet Cortical Stimulator 5 2 Revised 8 27 13 Blank Page...
Page 66: ...Nicolet Cortical Stimulator 5 4 Revised 8 27 13 Blank page...