10.
Set Stim
- At startup, the stimulation current level is set to 0 mA. Press the
Set Stim
button and then select the desired stimulation current level:
0 - 15
mA
(8 mA or less is typical) using the
Selector
knob.
NOTE:
If
Ele
or
Electrode -
are changed, the stimulation current level
will reset to 0 mA.
Begin with the lowest current level and increase in small increments.
11.
Mark Channel
- If you want to mark the channels with ID tags prior to and
after stimulation, press the
Mark
Channel
button.
12.
Check Stim
- Press the
Check
Stim
button to check the integrity of the
stimulator without stimulating the patient.
NOTE:
The maximum output voltage available is ±24V.
NOTE:
No precautions are required to be taken for a DC component in the
output since the output is AC coupled.
Summary of Contents for Nicolet SCU
Page 2: ......
Page 9: ...Revised 8 27 13 1 1 1 General Information...
Page 10: ...Nicolet Cortical Stimulator 1 2 Revised 8 27 13 Blank Page...
Page 21: ...Revised 8 27 13 2 1 2 Nicolet Cortical Stimulator Overview...
Page 22: ...Nicolet Cortical Stimulator 2 2 Revised 8 27 13 Blank Page...
Page 37: ...Revised 8 27 13 3 1 3 Performing a Cortical Stimulation Study...
Page 38: ...Nicolet Cortical Stimulator 3 2 Revised 8 27 13 Blank Page...
Page 55: ...Revised 8 27 13 4 1 4 Electromagnetic Compatibility EMC...
Page 56: ...Nicolet Cortical Stimulator 4 2 Revised 8 27 13 Blank Page...
Page 62: ...Nicolet Cortical Stimulator 4 8 Revised 8 27 13 Blank page...
Page 63: ...Revised 8 27 13 5 1 5 Error Codes...
Page 64: ...Nicolet Cortical Stimulator 5 2 Revised 8 27 13 Blank Page...
Page 66: ...Nicolet Cortical Stimulator 5 4 Revised 8 27 13 Blank page...