Revised 8/27/13
3-15
Performing a Cortical Stimulation Study
Using the Nicolet system to set the Cortical Stimulation
parameters
IMPORTANT: If you are currently stimulating the patient and you would
like to change any parameters, press the Stop button prior to making any
changes other than the Stimulation level (Set Stim parameter).
1. With the Nicolet system running in the Recorder mode, click on
View >
Panel > Cortical Stim
to display the Cortical Stim panel at the right side of
the screen display.
2.
Setup
- Click on the
Setup
show menu button and then click on
Probe
or
click on
Electrode
if using the Stimulus Switching Unit (SSU).
3.
Train Duration
- Click on the
Train Duration
button and then select the
desired train duration: Units of
seconds
(5 seconds, typical) or
Continuous
if using a stimulator probe.
4.
Positive Electrode
s - Click on the
‘Ele’
(Anode) show menu button
and then click on the desired
electrode number or label
.
5.
Negative Electrode
s - Click on the
‘Electrode -’
(Cathode) show menu
button and then click on the desired electrode number or label.
The applicable stimulation probe is the Natus Neurology Incorporated
Bipolar Stimulation Probe with 2.2mm diameter tips. The applicable
stimulation electrodes are any commercially available 2.3mm diameter
Strip and Grid Electrodes. In the United States only, the stimulation
electrodes and stimulation probes must have been cleared to market by
the Food and Drug Administration (FDA).
Summary of Contents for Nicolet SCU
Page 2: ......
Page 9: ...Revised 8 27 13 1 1 1 General Information...
Page 10: ...Nicolet Cortical Stimulator 1 2 Revised 8 27 13 Blank Page...
Page 21: ...Revised 8 27 13 2 1 2 Nicolet Cortical Stimulator Overview...
Page 22: ...Nicolet Cortical Stimulator 2 2 Revised 8 27 13 Blank Page...
Page 37: ...Revised 8 27 13 3 1 3 Performing a Cortical Stimulation Study...
Page 38: ...Nicolet Cortical Stimulator 3 2 Revised 8 27 13 Blank Page...
Page 55: ...Revised 8 27 13 4 1 4 Electromagnetic Compatibility EMC...
Page 56: ...Nicolet Cortical Stimulator 4 2 Revised 8 27 13 Blank Page...
Page 62: ...Nicolet Cortical Stimulator 4 8 Revised 8 27 13 Blank page...
Page 63: ...Revised 8 27 13 5 1 5 Error Codes...
Page 64: ...Nicolet Cortical Stimulator 5 2 Revised 8 27 13 Blank Page...
Page 66: ...Nicolet Cortical Stimulator 5 4 Revised 8 27 13 Blank page...